PREVENTIVE MAINTENANCE
a. Wash cabinet filter weekly.
b. The fine intake filter should be replaced
annually or every 2000 hours of use. More
frequently dusty /dirty environments.
c. Check oxygen concentration every 15,000
hours or 3 years of operation to verify the
continuing OCSI function.
Use original parts only
Contact Supera Anesthesia Innovations
for additiional information
877-620-1500
Electrical power supply:
115 V Units 230 V Units
Frequency:
60Hz
50/60Hz
Average Power:
490 watts
490/585 watts
Protection Class:
Class II
Class II
Mains Protection:
10A
5A
Filters:
At the rear of the device: a cabinet air filter.
At the compressor input: an inlet air filter, 5 µm,
located behind the cabinet air filter.
Air circulation
One tubeaxial fan cools the compressor compart
-
ment and a second fan cools the heat exchanger
coil.
.
Environmental limit conditions
The performances of the device (especially the
oxygen concentration) are quoted at 21
o
C (70
o
F)
and one atmosphere. They may change with
temperature and altitude.
• The device must be stored, transported and
used in the vertical position only.
• Ambient temperature of between 5
o
C and
40
o
C (40
o
F
to 104
o
F) operation.
• Storage temperature from -20
o
C to 60
o
C
(-4
o
F
to 140
o
F).
• Relative humidity of between 15% and 95%
operation and storage, both non-condensing.
• Altitude(21
o
C): Up to 2,286m (7,500ft)
without degradation; Consult your equip-
ment provider for further information re-
garding 2,286m to 4000m (7500 to 13000
ft)
• Complies with EN60601-1 standard; spil-
ling a glass of water.
5.7. Standards
EN 60601-1[UL60601-1:2003],
CAN/CSA-C22.2 No.601.1-M90 w/A1&A2:
Electrical Safety- Medical Devices.
EN60601-1-2:2001
Electromagnetic Compatibility