MAINTENANCE
Anesthesia Machine
The Supera Anesthesia Innovations anesthesia machine should be serviced by a
certified technician. Only certified service companies will have the factory specifi
-
cations and access to original replacement parts.
Annual Preventative Maintenance Service
Service will include:
Pre–service low and high pressure test. Inspection and replacement of hoses,
gaskets and seals as needed. Measure pressure relief valve resistance. Inspect
unidirectional valves. Inspection of evacuation system. Inspection of oxygen
system. Verification of vaporizer concentration output. Final low and high pres
-
sure test.
Contact Supera for a list of certified service companies.
Daily Inspection - Leak Test
Before each use check for pressure leaks in the machine. Also make certain the
waste gas evacuation system is working properly.
Oxygen Concentrator
*See the operators manual that came with the concentrator.
Your machine includes an oxygen concentrator and the filters must be kept clean
for proper operation.
There are two filters to monitor.
1. Coarse filter on the back of the machine. This can be washed with
mild soap and water. DO NOT INSTALL UNTIL 100% DRY
2. Fine particle filter under the course filter. Replace during the annual
PMS of the anesthesia machine. This is a replacement only filter.
Contact Supera for replacement. 877-620-1500
Failure to monitor the condition of the filters will void the warranty.