PREVENTIVE MAINTENANCE
a. Wash cabinet filter weekly.
b. Replace filter every 2000 hours or more often depending on environment.
c. Check oxygen concentration every 15,000 hours or 3 years of operation to verify the
continuing OCSI function.
Use original parts only
Contact Supera Anesthesia Innovations for additiional information
877-620-1500
Electrical power supply:
Rating:
115/230V 60Hz
230V 50Hz
Average Power:
330W(avg)
300 W(avg)
Protection Class:
Class II
Class II
Mains Protection:
5A
5A
Filters:
At the rear of the device: a cabinet air filter.
At the compressor input: an inlet air filter, 5 µm, located behind
the cabinet air filter.
Air circulation
A tubeaxial fan cools the compressor compartment.
Environmental limit conditions
The performances of the device (especially the oxygen concentra-
tion) are quoted at 21
o
C (70
o
F) and one atmosphere. They may
change with temperature and altitude.
• The device must be stored, transported and used in the vertical
position only.
• Ambient temperature of between 5
o
C and 40
o
C (40
o
F to 104
o
F)
operation.
• Storage temperature from -20
o
C to 60
o
C (-4
o
F to 140
o
F).
• Relative humidity of between 15% and 95% operation and
storage, both non-condensing.
• Altitude(21
o
C): Up to 2,286m (7,500ft) without degradation;
Consult your equipment provider for further information
regarding 2,286m to 4000m (7500 to 13000 ft)
• Complies with EN60601-1 standard; spilling a glass of water.
5.7.
Standards
EN 60601-1[UL60601-1:2003],
CAN/CSA-C22.2 No.601.1-M90 w/A1&A2:
Electrical Safety- Medical Devices.
EN60601-1-2:2001
Electromagnetic Compatibility