3.1 Turning on device
a. Plug the power cable into a power outlet
(Fig. 3.3) of the correct voltage and frequency
as defined on the manufacturer's technical
label (Fig 2.2).
b. Press the power switch
(I/
O
)
(item 1 in
Fig.2.1)
to the ON position
(I)
. The green
indicator light flashes until concentration is
achieved.
IMPORTAINT!
TURN THE ANESTHESIA MACHINES
OXYGEN FLOW METER ON TO A
MINUMUM FLOW RATE OF 0.25 LPM
A “NO FLOW” ALARM MAY SOUND IF
THE FLOW METER ISN’T OPEN
3.2 Turning off device
At the end of the usage, press the
I/
O
Switch
to place it in the
O
(OFF) position to stop the
device. The oxygen enriched air flow contin
-
ues for approximately one minute after the
device is stopped.
1. Filter / Silencer
2. Cabinet filter
3. Ventilation grill
Note: Shown with grate removed
Fig. 4.1
4. CLEANING - MAINTENANCE
Only the outside of the concentrator is to
be
cleaned. Use a damp sponge or cloth
with water only.
Acetone, solvents or any other inflammable
products must not be used.
Do not use abrasive powders.
FILTERS - IMPORTAINT!
Fine Filter (#1 fig. 4.1)
and Cabinet Filter Cleaning (#2)
The cabinet
filter (#2 Fig.4.1)
must be
cleaned in warm water and household
detergent weekly or after approximately
100 hours of use. Dry before reinstalling.
More frequent cleaning is recommended
in dusty environments.
Cabinet filter: p/n OC4000-1
The fine intake filter should be replaced
annually or every 2000 hours of use. More
frequently dusty /dirty environments.
Fine filter p/n OC4000-2