30
Maintaining and Cleaning the Oscar 2
6. Using the bulb of the mercury column, inflate the cuff to 250 mmHg. Once the pressure has
stabilized at this level, the LCD should match the mercury column by ±2.0 mmHg.
7. Check the unit against the column every 50 mmHg from 250 to 50 mmHg and the unit should
be within ±2.0 mmHg. If not, the unit needs to be returned to the service department for
recalibration or repair.
The Oscar 2 does not contain any user serviceable internal parts and should only be opened by an
authorized service representative. To return for service, send to your nearest SunTech office listed
under