Introduction to the Oscar Express ABP Monitor: Oscar 2
Specifications
Method of Measurement:
Oscillometry with step deflation.
Blood Pressure Range:
25-260 mmHg (max inflate 280mmHg)
Heart Rate Range:
40-200 bpm
Accuracy:
±3 mmHg
International Standards:
EN 60601-1, EN 60601-2-30,
EN 60601-1-2 (EMC),
EN 1060-1, EN 1060-3, "Non-Invasive
Sphygmomanometers - General
Requirements & Supplementary
Requirements for Electro-Mechanical BP
Measuring Systems”, AAMI SP10 ES1
category C’ (battery powered)
Operating Conditions:
10°C (50°F) to 50°C (122°F)
20-95% RH non-condensing
Power:
Two “AA” alkaline batteries or high
capacity rechargeable batteries (NiMH)
Calibration:
Minimally, once every year
Safety Systems:
Maximum inflation pressure limited to 300
mmHg; Auto safety release valve for power
failure. Maximum BP measurement time
limited to less than 180 seconds.
Size:
Approximately 120 x 80 x 32 mm
Weight:
Approximately 280g (including batteries)
Storage Conditions:
-20°c to 70°c, 15%-95% RH non-
condensing
Data Connector:
6 pin minidin:
1=PC-TXD (data from PC)
2= N/C
3= Ground
4= PC-RTS
5= PC-RXD (data to PC)
6= PC-DTR
Max voltage on any pin = ±15V with
respect to ground.