Troubleshooting

Overview of Quality and Test Codes

The columns labeled 

Test Code 

and

 Quality Code 

are used extensively in your data editing

decisions. These codes indicate various technical attributes of the BP reading that you should take
into consideration when making a decision to "tag" or "untag" a BP reading. It is possible to have
both a 

Test Code and a Quality Code 

on the same BP reading.

Remember, Test and Quality Codes do not automatically indicate that a reading is invalid;
they only serve as guides to help you edit data.

Oscar 2 test code definitions

Test
Code

Description in AccuWin Pro 

Solution

1

Weak or no oscillometric signal

Check position of cuff, tighten cuff.

2

Artifact/Erratic Oscillometric Signal

Remain still during BP reading.

3

Exceeded retry count (4 inflate
attempts)

Remain still during BP reading.

4

Exceeded Measurement time 

Check air hose connections and make
certain cuff is tight.

85

Reading Aborted (blocked values or
pneumatics)

Check air hose connections and make
certain air tubing is not crimped.

86

Reading Aborted (user abort)

Push START/STOP button to restart
reading.

87

Reading Aborted (inflate time-out out
or air leak)

Check air hose and cuff.

88

Reading Aborted (safety time-out)

Retry reading, push START/STOP button. If
problems persist return unit for servicing.

89

Reading Aborted (cuff over-pressure)

Check for blocked or kinked air hose.

90

Service Required (power supply out-
of-range or other hardware problem)

Replace batteries. If problem persists, return
for servicing.

91

Service Required (safety override
fitted or auto-zero out-of-range)

Retry by pushing START/STOP button. If
problems persist return unit for servicing. 

97 

Service Required Transducer out-of-
range

Return for servicing 

98

Service Required (A/D out-of-range)

Return for servicing

99

Service Required (EEPROM
calibration data CRC failure)

Unit needs to be recalibrated. Return for
servicing.

Summary of Contents for Oscar Express Remote Site

Page 1: ...Oscar Express Remote Site User s Guide...

Page 2: ...2...

Page 3: ...n 13 Specifications 14 Safety and Effectiveness 15 Warnings and Contraindications 15 Oscar 2 at a Glance 17 Preparing the Patient for Oscar 2 ABP Study 18 Oscar Express Data Transfer with Central Term...

Page 4: ...ns with diagnosis and management of hypertension The data obtained from ambulatory blood pressure monitors is highly accurate and useful for managing a wide variety of hypertensive situations includin...

Page 5: ...smission This eliminates the time and cost of mailing or transporting an ABP monitor between sites for remote data collection and ABP report generation Note Oscar Express is currently validated for us...

Page 6: ...ata transmission is an available analog phone line The remote site manages the clinical process of data collection issuing the monitor to patients setting up the patients with the monitor and making s...

Page 7: ...f items Your central terminal should have already set up your Oscar 2 monitor and programmed it for an ABP study thus making it ready for an immediate patient study To verify this turn on the monitor...

Page 8: ...ient ABP data and re program for the next study the proper connections need to be made in order to successfully perform a modem data transfer 1 Plug Oscar 2 to modem cable into the 25 pin port on the...

Page 9: ...ine as well a dual jack adapter is included to convert your wall jack into 2 jacks 4 Connect modem power cable to the modem and plug into to wall outlet The modem is now connected The final connection...

Page 10: ...ta transfer with the remote site the central terminal operator needs to communicate with the remote site to Identify which remote site is about to send data Identify the remote user they are speaking...

Page 11: ...ine connected to the modem is for data only The remote site makes a person to person call to the central terminal operator with the voice line and exchanges necessary information before performing a d...

Page 12: ...ssary information to perform the data transfer When instructed by the central terminal the remote site begins the data transfer over the same phone line Once modem communication is complete the modem...

Page 13: ...ed by the oscillometric method that senses pressure waves in the artery when occluded by pressure in the cuff Measurement of the frequency of the pressure waves also allows heart rate to be measured T...

Page 14: ...battery powered Operating Conditions 10 C 50 F to 50 C 122 F 20 95 RH non condensing Power Two AA alkaline batteries or high capacity rechargeable batteries NiMH Calibration Minimally once every year...

Page 15: ...not been established SunTech Medical Instruments Inc is responsible for the safety reliability and performance of its equipment under the following conditions Assembly operations extensions re adjust...

Page 16: ...ed for analog phone lines ONLY Data transfers through office PBX or other digital phone systems are not expected to be compatible with the Express modem data transfer Precautions for Use Ensure pressu...

Page 17: ...the remote site to the central terminal and can be programmed to automatically dial the central terminal Oscar Express enabled An Oscar 2 monitor displays an OE upon powering on to indicate the monit...

Page 18: ...rinter Indicates the number of readings in memory Percentage Indicates what percentage of Express data transfer is complete Preparing the Patient for Oscar 2 ABP Study In order to accurately measure b...

Page 19: ...the Oscar 2 into its pouch with the display showing through the window Fig 5 Attach the pouch to the patient using either the shoulder strap or belt depending on the patient s preference Fig 6 Ensure...

Page 20: ...uff sleeve is directly over the brachial artery Fit the cuff snugly around the upper arm While sleeping the patient should make sure that the hose is not crimped The batteries can be replaced during a...

Page 21: ...ode turn the monitor off and then power the monitor back on The first step in performing a data transfer is to call a central terminal operator to identify your remote site indicate you have patient d...

Page 22: ...hile the central terminal operator can communicate to you the status of the data transfer over the voice line Step 4 Completing communication with Central Terminal When the data transfer is complete t...

Page 23: ...Warning Any background noise during the data transfer can interfere with the modem transmission If the remote site phone has a mute button it should be used to minimize background noise from interferi...

Page 24: ...protocol at any time and initiate a reading if you have programmed the Oscar 2 monitor to do so Indicates that the pressure to which the cuff inflated in the dynamic inflation mode may have been too...

Page 25: ...ial does not work or dials wrong number Use phone to manually dial central terminal Central terminal should check the remote site information window for correct auto dial number settings If the Oscar...

Page 26: ...ement time Check air hose connections and make certain cuff is tight 85 Reading Aborted blocked values or pneumatics Check air hose connections and make certain air tubing is not crimped 86 Reading Ab...

Page 27: ...that would indicate that this reading is suspect Refer to the table above for detailed definitions of Test Codes LOH Dynamic pressure too low Pressure to which cuff inflated in the dynamic inflation...

Page 28: ...dian and Latin American Headquarters and Service 507 Airport Boulevard Suite 117 Morrisville NC 27560 8200 Toll free USA and Canada 800 421 8626 Phone 919 654 2300 Fax 919 654 2301 www SunTechMed com...

Page 29: ...lly inspect cables pneumatic hoses and the monitor case for cracks fraying or kinks DO NOT use the monitor if it shows any signs of damage Contact the SunTech service department Annual Maintenance It...

Page 30: ...k the unit against the column every 50 mmHg from 250 to 50 mmHg and the unit should be within 2 0 mmHg If not the unit needs to be returned to the service department for recalibration or repair The Os...

Page 31: ...98 0063 01 0 1 Orbit ABP cuff Adult 98 0063 02 1 1 Orbit ABP cuff Large Adult 98 0063 03 0 1 Oscar 2 Pouch 98 0032 00 1 1 Oscar 2 Strap 98 0036 00 1 1 Oscar 2 Belt 98 0037 00 1 1 Patient Diaries 82 00...

Page 32: ...cted defect The instrument should be carefully packaged and shipped prepaid to SunTech Medical 507 Airport Boulevard Suite 117 Morrisville NC 27560 8200 Tel 1 800 421 8626 or 1 919 654 2300 Fax 1 919...

Page 33: ...33 80 0031 00 Part 80 0031 00 Rev A...

Reviews: