26
Troubleshooting
Overview of Quality and Test Codes
The columns labeled
Test Code
and
Quality Code
are used extensively in your data editing
decisions. These codes indicate various technical attributes of the BP reading that you should take
into consideration when making a decision to "tag" or "untag" a BP reading. It is possible to have
both a
Test Code and a Quality Code
on the same BP reading.
Remember, Test and Quality Codes do not automatically indicate that a reading is invalid;
they only serve as guides to help you edit data.
Oscar 2 test code definitions
Test
Code
Description in AccuWin Pro
Solution
1
Weak or no oscillometric signal
Check position of cuff, tighten cuff.
2
Artifact/Erratic Oscillometric Signal
Remain still during BP reading.
3
Exceeded retry count (4 inflate
attempts)
Remain still during BP reading.
4
Exceeded Measurement time
Check air hose connections and make
certain cuff is tight.
85
Reading Aborted (blocked values or
pneumatics)
Check air hose connections and make
certain air tubing is not crimped.
86
Reading Aborted (user abort)
Push START/STOP button to restart
reading.
87
Reading Aborted (inflate time-out out
or air leak)
Check air hose and cuff.
88
Reading Aborted (safety time-out)
Retry reading, push START/STOP button. If
problems persist return unit for servicing.
89
Reading Aborted (cuff over-pressure)
Check for blocked or kinked air hose.
90
Service Required (power supply out-
of-range or other hardware problem)
Replace batteries. If problem persists, return
for servicing.
91
Service Required (safety override
fitted or auto-zero out-of-range)
Retry by pushing START/STOP button. If
problems persist return unit for servicing.
97
Service Required Transducer out-of-
range
Return for servicing
98
Service Required (A/D out-of-range)
Return for servicing
99
Service Required (EEPROM
calibration data CRC failure)
Unit needs to be recalibrated. Return for
servicing.