Table 2  Troubleshooting

* If you have already tried the above solutions and the device still cannot nebulize normally, we 

suggest you to contact your place of purchase.

Unsure which 

medicine is suitable 

for nebulization

Water buildup in 

mouthpiece 

obstructs mist

Indicator light 

is blue

Condensation

Remove the mouthpiece and pour 

out all liquid.

Single patient use

It is necessary to avoid 

cross infection.

Everyone must use their 

Accessories only by themselves.

Liquid medicine in 

the Medicine Cup 

is leaking

Medicine Cup is broken or 

plastic is worn out.

Replace with new Medicine Cup.

Noise becomes 

louder towards end 

of nebulization cycle

The mesh plate vibrates 

without load while the 

liquid is running out.

Normal phenomenon.

Can not turn on/off 

the device

1. The lithium battery has not 

been charged sufficiently.

2. The device gets stuck.

1. Recharge battery before using.

2. Press the reset button beside the 

USB interface on the bottom of the 

unit with a small pointed object.

The device can not 

turn off by itself 

when the drug is 

exhausted

1. Some foam produced by drug 

is left in the cup.

2. Medicine Cup failure.

3. There is drug/water left or 

stains on the electrode. 

1. Press the Power button and clear 

the bubble.

2. Please contact the authorized 

service center or manufacturer.

3. Press the Power button and clear 

the electrode.

Consult your doctor.

The battery is low.

Recharge battery before using.

Only use under guidance of 

doctor's prescription.

The device is on 

but weak

1. Recharge the battery before use

2. Clean the device by adding water to 

the Medicine Cup, press the power 

button for 3 seconds, then let the 

self-cleaning cycle run until complete.

3. The liquid medicine is unfit for this 

equipment, please consult a doctor.

4. Please contact the authorized service 

center or manufacturer

5. Clean the contact points before use

PROBLEM

No mist

CAUSE

SOLUTION

1. The battery is very low

2. The equipment is damaged

3. There are stains on metal 

contact points of nebulizer in 

contact with the Medicine Cup

1. The device and Medicine Cup 

are not correctly assembled.

2. The drug produces too much 

foam.

3. The mesh plate is clogging.

4. The equipment is damaged.

1. Battery has low electricity.

2. The mesh plate is clogging.

3. The medicine is not fit for this 

nebulizer.

4. The mesh plate in the 

Medicine Cup is damaged.

5 .The metal contacting point 

between Medicine Cup and 

main unit is not clean.

1. Correct the device by pressing back 

button, releasing, and introduce the 

Medicine Cup once again until you 

hear a "click".

2. Shake the device slightly or stand 

the device for a while until the 

foam disappears and then restart 

the nebulizer manually.

3. Clean the device by adding water to 

the Medicine Cup, press the power 

button for 3 seconds, then let the 

self-cleaning cycle run until 

complete.

4. Contact your place of purchase.

1. Recharge the battery before use

2. Contact your place of purchase

3. Clean the metal contact points of 

nebulizer and try again

PROBLEM

CAUSE

SOLUTION

20

21

Chapter 6   Troubleshooting

Cannot turn on

Summary of Contents for NEB401

Page 1: ...uction Manual Mesh Nebulizer with Rechargeable Battery NEB401 Rx Only Manufactured for Sunset Healthcare Solutions 180 N Michigan Ave Ste 2000 Chicago IL 60601 www sunsethcs com NEB401 IFU V1 01 2020...

Page 2: ...nt the device may be used with pediatric 4 years of age and adult patients in the home hospital and sub acute care settings It is not intended for use with Pentamidine Table of Contents Chapter 1 Befo...

Page 3: ...ooting Chapter 7 Customer Service Electromagnetic Compatibility Notes on FCC certification Chapter 9 Electromagnetic Compatibility Symbol Glossary Chapter 10 Symbol Glossary Cleaning Disinfecting Clea...

Page 4: ...with pediatric 4 years of age and adult patients in the home hospital and sub acute care settings It is not intended for use with Pentamidine 1 What is the Mesh Nebulizer with Rechargeable Battery 2...

Page 5: ...nvironment Warning For type or dose of medicine please follow the instructions of your physician and or licensed healthcare practitioner Important Safety Information Read Before Use The Mesh Nebulizer...

Page 6: ...on in the treatment area such as sores herpes simplex or open wounds If you are not sure the device is safe for you or not please consult your doctor PRECAUTIONS Take the following precautions to avoi...

Page 7: ...e end of USB cord to an adapter or other USB port such as USB port on a computer This nebulizer is powered with a built in lithium battery The battery can be charged with the USB cord 1 Power Supply C...

Page 8: ...ed in the right place the device may not be able to operate normally NOTE Picture 5 Mask Assembly and Mouthpiece Assembly 4 Usage of Nebulizer Nebulizer Accessories are categorized as mouthpiece adult...

Page 9: ...hile nebulizing 1 If this nebulizer is being used to treat a life threatening condition a backup device is necessary 2 The nebulizer may be unable to be used due to power failure or mechanical malfunc...

Page 10: ...n off the power The Power Indicator Green Light will go out 1 Cleaning Chapter 5 Maintenance of the Nebulizer Be sure to turn off the device After each use first discard all residual medicine and remo...

Page 11: ...r 3 6 ml 1 5 Press and hold the Power button for 3 seconds WHILE THE DEVICE IS OFF this will turn on the light and alternate between green and blue colors Release the button allowing the device to ent...

Page 12: ...utes for each time 3 years for main unit 3 months for Medicine Cup For the battery the device can be used for approximately 8 days after charging if operating for 30 minutes a day Lithium battery has...

Page 13: ...se under guidance of doctor s prescription The device is on but weak 1 Recharge the battery before use 2 Clean the device by adding water to the Medicine Cup press the power button for 3 seconds then...

Page 14: ...red equipment Nebulization Rate 0 2ml min minimum Particle Size 2 2 m 25 Medicine Capacity 8ml maximum 0 5ml minimum Battery Life the battery can be used for approximately 8 days after charging if ope...

Page 15: ...Warning Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equipme...

Page 16: ...issions are very low and are not likely to cause any interference in nearby electronic equipment The Mesh Nebulizer with Rechargeable Battery is suitable for use in all establishments including domest...

Page 17: ...ectromagnetic environment guidance 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV for power supply lines 1 kV differential mode 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 7...

Page 18: ...fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked wit...

Page 19: ...hat to which the receiver is connected Consult the dealer or an experienced for help This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This devi...

Page 20: ...Name Tel Sales shop name Sales shop address Address Zip code Product model Date of purchase Dealer info Mesh Nebulizer with Rechargeable Battery Service Warranty Card...