12
13
Picture 7
(1) Please use mouthpiece correctly by tightening, and make sure
liquid medication does not go directly to the mouth.
(2) Slightly tilt the Main Unit toward yourself to immerse the
Vibrating Mesh in the medicine as the supply diminishes.
(3) If the Low Battery Light (Blue) continues to flash
slowly,it means the power is running out, please
charge device as soon as possible.
(4) The dosage of medication must follow
physician's prescribed treatment.
(5) Do not press the two single-way switches
(inhalation switch and exhalation switch)
on Medicine Cup, which may cause
failure of the mesh technology.
Children must be under adult supervision while nebulizing.
(1) If this nebulizer is being used to treat a life threatening condition, a backup device
is necessary.
(2) The nebulizer may be unable to be used due to power failure or mechanical
malfunction. In order to avoid emergencies such as accidents or acute asthma
attacks, it is recommended to have a backup device.
Do not force
NOTE !
NOTE !
CAUTION !
Picture 6 start to aerosolize
(3) Sitting upright, calm, and relaxed, hold the nebulizer carefully. Take deep and slow
breaths for a more effective, enjoyable treatment.
Press
Foam is likely to be produced and accumulate near the mesh during operation,
which may lead to a decrease of aerosol output and the nebulizer shutting
down automatically. In this situation please shake the device slightly or stand
the device for a while until the foam disappears and then restart the nebulizer
manually.
NOTE !
Summary of Contents for NEB401
Page 21: ...WARNING No modifications of this equipment are allowed...
Page 22: ......