background image

LIT-239 

SUNOPTIC SURGICAL

®

                                                                                  

Rev. A

                                                                                                                                                                                  

(English)

                                                                                                              

Page 8 of 21 

 

 
 
 
 
 
 
 
 

 

Figure 2 - Camera Console Rear Panel

 

 
 
 
 

No. 

Name 

Function 

Fan 

Provides air flow for cooling of unit 

Grounding Connector 

For electric potential equalization 

Camera Port 

Accepts the optional second camera cable 

Circuit Breakers 

Over current protection 

AC Power Entry and Switch  Accepts AC Power cord and switches on AC power 

Video Output Ports 

Two HDMI and two DVI ports provide video output to monitor 

USB Input Ports 

Port functions from left to right 
1: Keyboard Port 
2: No Function 
3 and 4: Used for firmware updates, not to be used for 
keyboard or file transfer

.

 

Label 

Regulatory Product Label 

Recorder Ethernet Port 

Allows user to connect to recorder via Ethernet 

10 

Stabilizer Ethernet Port 

Allows user to connect to stabilizer via Ethernet 

11 

Foot Pedal Port 

Accepts the 2.5mm plug from foot pedal 

 

 
 
 
 
 

10 

11 

Summary of Contents for HDC-300C

Page 1: ...n HD Camera Instructions for Use Version 1 0 Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville FL 32216 USA Customer Service 904 737 7611 Toll Free 877 677 2832 EC REP AJW Technology Consultin...

Page 2: ...SETTING UP THE CONSOLE 4 2 CONNECTING THE CAMERA 4 3 CONNECTING MONITORS 4 4 CONNECTING AC POWER 4 5 POWERING ON THE UNIT 4 6 TURNING OFF THE UNIT 5 OPERATION 5 1 USING THE CONSOLE TOUCH SCREEN INTER...

Page 3: ...tor treat or alleviate disease injuries or handicaps nor for the investigation replacement or modification of the anatomy or of a physiological process and is not to be used with or as an endoscope or...

Page 4: ...o 50 C 10 to 90 non condensing Operating Conditions 50 to 86 F 10 to 30 C 30 o 75 RH non condensing 1 4 CLASSIFICATIONS PARAMETER VALUE System Classification FDA Class I EU Class I Health Canada Class...

Page 5: ...lete knowledge of the use of this equipment and the procedure to be performed 3 This equipment should be installed and tested prior to use A pre operative check should be performed prior to administra...

Page 6: ...nergy which may affect the normal function of nearby installed equipment Equipment that generates and radiates RF energy may affect the normal function of the HD Camera System When choosing a location...

Page 7: ...CAMERA CONSOLE Figure 1 Camera Console Front Panel No Name Function 1 Touch Screen Controls all aspects of Camera Console 2 On Button Turns the unit on or off 3 USB Storage Port For external flash dr...

Page 8: ...Power cord and switches on AC power 6 Video Output Ports Two HDMI and two DVI ports provide video output to monitor 7 USB Input Ports Port functions from left to right 1 Keyboard Port 2 No Function 3...

Page 9: ...ight source 2 Light Guide Project fiber optic light onto surgery site 3 Camera Lens Lens through which video is recorded 3 3 COMPATIBLE ACCESSORIES Figure 4 Compatible Accessories for the Camera Syste...

Page 10: ...e the unit 4 2 CONNECTING THE CAMERA WARNING THE CAMERA CORD HAS A LOCKING CONNECTOR TO REMOVE THE CABLE FROM THE UNIT GRAB THE CONNECTOR AND PULL THE CONNECTOR OUT THE CONNECTOR CANNOT BE REMOVED BY...

Page 11: ...transformers and or power strips only CAUTION Nearby radio or TV units may subject this equipment to radio interference To avoid adverse electromagnetic effects DO NOT operate this equipment near any...

Page 12: ...nd finishes its boot sequence the Home Screen will appear The following options are available 5 2 ENTER PATIENT DATA This menu option will detail instructions on how to access the Patient Data entry f...

Page 13: ...eo recording if patient data is entered Case will be stored with name otherwise Case will be automatically named with the date and time video starts Pause Pauses video recording pressing this button a...

Page 14: ...e language on the recording device Photo Quality Select desired quality for pictures Format Select desired format for pictures Video Quality Select desired quality for videos EQ Economy Quality lowest...

Page 15: ...when the unit is started Folders will be erased starting with the oldest data to keep either 200GB 100GB 50GB the last 24 hours of data or the last 7 days of data Select the OFF option to avoid automa...

Page 16: ...s associated with it For a file with multiple indices OR videos and pictures use the up and down arrows to access the additional files Press F6 or F12 or access Record Close File and Exit to exit play...

Page 17: ...d playback and to take pictures without accessing the keyboard or the touch screen Plug the footswitch cable into the 2 5mm jack on the back of the unit Figure 5 TASCAM Footswitch No Name Function 1 S...

Page 18: ...ntifying the fault and how to correct it FAN FAULT In the event of a fan problem the unit will display this screen Turn off and restart the Camera System If the fault persists the unit needs servicing...

Page 19: ...nary ammonium compounds or hydrogen peroxide The lenses should be cleaned only with lens tissue available in any camera store follow the directions on the package Do not use strongly caustic or acidic...

Page 20: ...icing under this warranty please contact Sunoptic Technologies for return authorization documentation Please carefully pack the unit in a sturdy carton and ship it to the factory Include a note descri...

Page 21: ...consult accompanying documents Caution Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner Consult Instructions for Use CE mark Not for disposal in gene...

Reviews: