11
Clearing Memory
After printing the report, clear the memory using the key press options in the
table below.
Checking Unit Status
Before starting a new test, it is recommended to scroll through the status items
to confirm they are correct.
–
Current 1.2
Displays the radon concentration in the current
12-hour period.
–
Start Test = 5
AVG or CUR reading is complete. The data for the
test remains in memory and must be cleared
before starting a new test. See
Press…
Display Shows…
Comment
–
Start Test = 5
It is recommended to print a report before clear-
ing data.
+
Prt=3 Cl=4 Ex=5
When the “
Start Test = 5
” message appears,
press and hold 3 and 4. Keep holding until the
“
Prt=3 Cl=4 Ex=5
” message appears. Press 4 to
clear (or 5 to exit).
Confirm Y=5 N=2
Press 5 to confirm the clear memory command.
Clearing Memory
The memory is cleared in a few seconds.
Press…
Display Shows…
Comment
–
Start Test = 5
The status menu can be selected when the display
shows “Start Test = 5.”
Status Menu
Press the (1) button to select the status menu. The
status of each item will scroll through automati-
cally. Each item displays for about 2 seconds.
–
DELAY nn hr
The delay time selected.
–
INTERVAL nn hr
The measurement interval selected.
–
DURATION nnn hr
The duration of the test.
–
BATTERY Ok
Battery status.
–
TIME nn:nn PM
Current time.
–
DATE MM DD YY
Current date
–
CAL MM/DD/YYYY
Date of last calibration.
–
SNC MODEL 1028
Model number.