BED Awareness System Error Codes
Error Handling
• The error handling system provides a mechanism for capturing errors known to the product and providing feedback
to the operator through logged errors or message errors.
• All Logged errors generate a generic blinking error message which will be displayed on the status screen in the
lower area: “Error XXX Call Service”, where XXX is the error code.
• Errors will only be logged if the error log does not already contain the error.
• In the event multiple errors are displayed on the status screen, the error code is used to identify the priority. The
lowest code has the highest priority.
• All logged errors and message errors are persistent until the power is cycled, or the condition is fixed.
• The details for each logged error and error message are detailed in the table below.
Logged Errors
Error
Code
Name
Description
Type
Affected Systems
100
Not Used
Highest Priority Errors
102
StuckSafetyRelay
This error is checked while the product is
“On”. If the safety relay signal is “On” and
there is no button press on the product then
set this error after 1 minute.” Note: This should
not be checked when in Auto-Cycle Mode.
Logged
Motion
103
InputTimer
If any footboard motion or side rail/pendant
input is sensed to be “On” continuously for a
total of 5 minutes then provide an exception
message. Note: This should not be checked
when in Auto-Cycle Mode.
Logged
Motion
200
Not Used
Scale System
201
HL-Load Cell
Patient Head Left Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
202
HR-Load Cell
Patient Head Right Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
203
FL-Load Cell
Patient Foot Left Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
204
FR-Load Cell
Patient Foot Right Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
206
CS5533Fatal
Non-Recoverable
Communication
Error.
CS5533 Fatal occurs when the CPU fails
to communicate to the External ADC for the
Load cells and cannot recover.
Logged
Scale, Bed Exit,
i
BED Awareness
207
ZeroFail
Zero Fail error occurs when the system can-
not provide a stable zero value for three con-
secutive attempts at zeroing the product.
Logged
Scale, Bed Exit,
i
BED Awareness
Summary of Contents for Zoom 2040
Page 2: ......
Page 38: ...38 2040 309 002 REV B www stryker com Notes Return To Table of Contents ...
Page 86: ...86 2040 309 002 REV B www stryker com Base Assembly and Options Return To Table of Contents ...
Page 107: ...www stryker com 2040 309 002 REV B 107 ZOOM Base Assembly Return To Table of Contents ...
Page 108: ...108 2040 309 002 REV B www stryker com ZOOM Base Assembly Return To Table of Contents ...
Page 118: ...118 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Page 119: ...www stryker com 2040 309 002 REV B 119 Litter Assembly Return To Table of Contents ...
Page 120: ...120 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Page 121: ...www stryker com 2040 309 002 REV B 121 Litter Assembly Return To Table of Contents ...
Page 127: ...www stryker com 2040 309 002 REV B 127 Litter Assembly Return To Table of Contents ...
Page 128: ...128 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Page 148: ...148 2040 309 002 REV B www stryker com ZOOM Litter Assembly Return To Table of Contents ...
Page 149: ...www stryker com 2040 309 002 REV B 149 ZOOM Litter Assembly Return To Table of Contents ...
Page 150: ...150 2040 309 002 REV B www stryker com ZOOM Litter Assembly Return To Table of Contents ...
Page 151: ...www stryker com 2040 309 002 REV B 151 ZOOM Litter Assembly Return To Table of Contents ...
Page 160: ...160 2040 309 002 REV B www stryker com Head End Siderail Assembly Return To Table of Contents ...
Page 176: ...176 2040 309 002 REV B www stryker com Foot End Siderail Assembly Return To Table of Contents ...
Page 178: ...178 2040 309 002 REV B www stryker com Foot End Siderail Assembly Return To Table of Contents ...