Problem / Failure
Recommended Action
No Gatch down motion.
A. Check for 5 VDC on TP 5 on the CPU board referencing
ground test point while pressing gatch down.
a. If 5 VDC is present, go to step B.
b. If 5 VDC is not present, replace CPU board.
B. Check for 120 VAC power on connector CC, pin 2 (red)
and pin 3 (white), of the CPU board while pressing
gatch down.
a. If voltage is not present, replace the CPU board.
b. If 120 VAC is present, check the capacitor and
motor.
C. Verify bed function and return to service.
No Gatch up motion.
A. Check for 5 VDC on TP 6 on the CPU board referencing
ground test point while pressing gatch up.
a. If 5 VDC is present, go to step B.
b. If 5 VDC is not present, replace CPU board.
B. Check for 120 VAC on connector CC, pin 1 (black)
and pin 3 (white), of the CPU board while pressing
gatch up.
a. If voltage is not present, replace the CPU board.
b. If 120 VAC is present, check the capacitor and
motor.
C. Verify bed function and return to service.
No Fowler up/or uneven motion.
A. Check for 5 VDC on TP 3 on the CPU board referencing
ground test point while pressing Fowler up.
a. If 5 VDC is present, go to step B.
b. If 5 VDC is not present, replace CPU board.
B. Check for 120 VAC on connector GG, Pin 1 (white)
and pin 2 (black), of the CPU board while pressing
Fowler up.
a. If voltage is not present, replace the CPU board.
b. If 120 VAC is present, check the capacitor and
motor.
C. Refer to Fowler Mechanism Customer Guide
(2030-009-028).
D. Verify bed function and return to service.
No Fowler down/or uneven motion.
A. Check for 5 VDC on TP 4 on the CPU board referencing
ground test point while pressing Fowler down.
a. If 5 VDC is present, go to step B.
b. If 5 VDC is not present, replace CPU board.
B. Check for 120 VAC on connector GG, Pin 1 (white) and
pin 3 (red), of the CPU board while pressing Fowler up.
a. If voltage is not present, replace the CPU board.
b. If 120 VAC is present, check the capacitor and
motor.
C. Refer to Fowler Mechanism Customer Guide
(2030-009-028).
D. Verify bed function and return to service.
Troubleshooting Guide
Summary of Contents for Zoom 2040
Page 2: ......
Page 38: ...38 2040 309 002 REV B www stryker com Notes Return To Table of Contents ...
Page 86: ...86 2040 309 002 REV B www stryker com Base Assembly and Options Return To Table of Contents ...
Page 107: ...www stryker com 2040 309 002 REV B 107 ZOOM Base Assembly Return To Table of Contents ...
Page 108: ...108 2040 309 002 REV B www stryker com ZOOM Base Assembly Return To Table of Contents ...
Page 118: ...118 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Page 119: ...www stryker com 2040 309 002 REV B 119 Litter Assembly Return To Table of Contents ...
Page 120: ...120 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Page 121: ...www stryker com 2040 309 002 REV B 121 Litter Assembly Return To Table of Contents ...
Page 127: ...www stryker com 2040 309 002 REV B 127 Litter Assembly Return To Table of Contents ...
Page 128: ...128 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Page 148: ...148 2040 309 002 REV B www stryker com ZOOM Litter Assembly Return To Table of Contents ...
Page 149: ...www stryker com 2040 309 002 REV B 149 ZOOM Litter Assembly Return To Table of Contents ...
Page 150: ...150 2040 309 002 REV B www stryker com ZOOM Litter Assembly Return To Table of Contents ...
Page 151: ...www stryker com 2040 309 002 REV B 151 ZOOM Litter Assembly Return To Table of Contents ...
Page 160: ...160 2040 309 002 REV B www stryker com Head End Siderail Assembly Return To Table of Contents ...
Page 176: ...176 2040 309 002 REV B www stryker com Foot End Siderail Assembly Return To Table of Contents ...
Page 178: ...178 2040 309 002 REV B www stryker com Foot End Siderail Assembly Return To Table of Contents ...