![Stryker Visum Operation Manual Download Page 30](http://html.mh-extra.com/html/stryker/visum/visum_operation-manual_1386072030.webp)
30
S
8. Optional Configurations
8.1 In-Light Camera (Camera-Ready Light 1 only)
8.1.1 Swapping Weighted Handle for the In-Light Camera
Note: An HD Camera Decoder box is required for HD camera installation.
If the light system is camera-ready, the light attached to the
uppermost arm segment of the suspension (Light 1) is the
one designed to hold the in-light camera. The other lights
are not camera-ready, so these steps apply to Light 1 only.
1. Invert the light head so that the weighted light handle
points straight up, perpendicular to the floor.
2. Loosen the three thumb screws securing the weighted
light handle to the light.
3. Pull the weighted light handle free from the light head.
In-Light Camera
Caution
When you remove the weighted light handle, the Spring Arm may drift up. Be prepared to hold it
down until the camera is secure.
4. Align the serial port connector of the In-Light Camera with that of the light head and insert it into the
light head.
5. Press the In-Light Camera firmly into place, and ensure that the base of the In-Light Camera sits flush
with the surface of the light.
6. Tighten the three thumb screws completely to secure camera.
Warning
If the base of the In-Light Camera is not flush with the surface of the light or if the thumb screws are
not completely tightened, the camera could fall off.