Sterilization**
Individually wrap in two layers of polypropylene wrap, (Kimguard KC600) or
similar.
Place in sterilizer with the following parameters:
Sterilization
Type
Preconditioning
Pulses
Minimum
Temperature
Full
Cycle
Time
Minimum
Dry Time
Gravity-
displacement
Steam Sterilization
– US (wrapped)
N/A
132°C
15
minutes
45
†
minutes
Gravity-
displacement
steam sterilization
– EU (wrapped)
N/A
134°C
10
minutes
45
†
minutes
Pre-vacuum steam
sterilization --US
(wrapped)
3
132°C
4
minutes
20
minutes
Pre-vacuum steam
sterilization --EU
(wrapped)
3
134°C
3
minutes
20
minutes
Storage
Keep in instrument tray until use.
** Sterilizable camera cover must be sterilized according to ISO17665-1.
†
The validated gravity dry times exceed the AAMI ST79 standard dry time range for wrapped devices by
15 minutes.
Note:
Any modifications to these parameters should be based on standard hospital
cleaning/sterilization protocol, and are performed at the discretion of the user, who assumes
the risk for modifying these parameters.