![Stryker Visum Operation Manual Download Page 22](http://html.mh-extra.com/html/stryker/visum/visum_operation-manual_1386072022.webp)
22
S
Sterilization**
Individually wrap in two layers of polypropylene wrap, (Kimguard KC600) or
similar.
Place in sterilizer with the following parameters:
Sterilization
Type
Preconditioning
Pulses
Minimum
Temperature
Full
Cycle
Time
Minimum
Dry Time
Gravity-
displacement
Steam Sterilization
– US (wrapped)
N/A
132°C
15
minutes
20
minutes
Gravity-
displacement
steam sterilization
– EU (wrapped)
N/A
134°C
10
minutes
20
minutes
Pre-vacuum steam
sterilization --US
(wrapped)
3
132°C
4
minutes
20
minutes
Pre-vacuum steam
sterilization --EU
(wrapped)
3
134°C
3
minutes
20
minutes
Storage
Keep in instrument tray until use.
** Sterilizable handle cover must be sterilized according to ISO17665-1.
Note:
Any modifications to these parameters should be based on standard hospital
cleaning/sterilization protocol, and are performed at the discretion of the user, who assumes
the risk for modifying these parameters.
Instructions for Sterilizable Camera Cover
Point of Use
Wipe excess soil from the sterilizable camera cover using disposable paper
towels.
Containment and
Transportation
Reprocess the sterilizable camera cover as soon as reasonably practical
following use.
Preparation
for Cleaning
1. Properly prepare an enzymatic detergent according to the detergent
manufacturer’s recommendations.
2. Dip a clean soft cloth into the detergent and wipe the entire surface of the
sterilizable camera cover for a minimum of 2 minutes each.
3. Immerse the sterilizable camera cover in the detergent.
4. Soak the sterilizable camera cover in the detergent for at least 15 minutes.