Figure 22
Instrument Bar
17
>
Once size confirmation is complete, attach the 1/8” Peg Drill to the
Universal Driver and create fixation pin holes (for the 4:1 Cutting
Block) through the holes on the face of the Femoral Sizer marked
“EPI” .
>
Remove the Headed Pins using the Headed Pin Extractor .
Note: With the use of the Single-Use 4:1 Cutting Block there is the
potential for debris generation . It is recommended that the saw blade be
fully inserted into the slot prior to initiating oscillation . Additionally, it is
recommended that an irrigation system be available and that the joint be
thoroughly irrigated after use of the block .
Femoral
Preparation
6541-5-500
MIS AP Sizer Adjustment Housing
6541-5-508
MIS AP Sizer Body - Left
6541-5-509
MIS AP Sizer Body - Right
6541-5-510
MIS Femoral Stylus
6541-4-802
1/8” Hex Drive
6541-4-518
1/8” Peg Drill
6541-4-801
Universal Driver
6541-4-300
Headed Pin Impactor Extractor
6541-4-515
Headed Nails - 1 1/2”
Figure 23
>
Locate the fixation pegs of the appropriate size Single-Use 4:1 Cutting
Block into the pin holes created on the distal femur .
See Catalog
Triathlon Single-Use Instruments
Femoral Prep Kit
Single-Use 4:1 Cutting Block