Troubleshooting Guide (cont’d)
Periodic Maintenance
Storage and Handling
To ensure the longevity, performance and safety of this equipment, use the original packaging materials when stor-
ing or transporting this equipment.
INTERVAL
ACTIVITY
Prior to each use.
Inspect, operate and test the handpiece to ensure that it is working properly. ensure that there are no
loose or missing components. Check all moving parts for free movement. Be alert for unusual sounds or
vibrations and note the operating speed.
PROBLEM
CAUSE
ACTION
Handpiece has become noisy
and vibrates.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Sporadic electrical interference
is experienced.
electrical noise is present.
Turn off all electrical equipment not in use in
the operating room.
Relocate electrical equipment; increase
spatial distance.
Plug operating room equipment into different
operating room outlets.