10
www.stryker.com
2008/01
6205-001-700 Rev-A
WARNINGS:
• Clean and sterilize handpieces, attachments and bat-
teries before first and every use.
• Prior to cleaning and sterilization, remove attach-
ments and all accessories from the handpieces.
• Orient the handpiece to allow steam to flow through
the cannulated handpiece when using a gravity dis-
placement sterilizer.
To obtain optimal performance and prevent damage,
perform one of the following sterilization procedures:
“Flash” Autoclave:
• Gravity displacement sterilizer
• 270-272 °f (12-1 °C)
• Unwrapped in an instrument tray in a vertical position
so steam can flow through the cannula
• 10-minute minimum exposure time
Hi Vac:
• Pre-vacuumed sterilizer
• 270-272 °f (12-1 °C)
• Wrapped or unwrapped
• -minute minimum exposure time
• 8-minute minimum dry time
To Sterilize Battery Packs
See the care instructions supplied with the battery
packs.
To Sterilize Handpieces and Attachments
Sterilization Recommendations*
ETO:
• 100% eTO
• 120-15 °f (-57 °C)
• Wrapped in an instrument tray or fully perforated
sterilization box
• 2-hour 0-minute exposure time, 8-hour minimum
aeration time
250 °F Gravity:
• Gravity displacement sterilizer
• 250-25 °f (121-12 °C)
• Wrapped in an instrument tray or fully perforated
sterilization box
• 50-minute exposure time
• 8-minute minimum drying time
270 °F Gravity:
• Gravity displacement sterilizer
• 270-272 °f (12-1 °C)
• Wrapped in an instrument tray or fully perforated
sterilization box
• 5-minute minimum exposure time
• 8-minute minimum dry time
*Validation is based on the Association for the Advancement of
Medical Instrumentation (AAMI) protocol.
NOTE:
After sterilization, allow the equipment to cool
to room temperature to ensure a comfortable operating
temperature.