
8
3006-209-001 REV B
www.stryker.com
Introduction
Expected service life
The 3002
S3
MedSurg bed has a ten year expected service life under normal use conditions and with appropriate
periodic maintenance.
The optional motion/nurse call pendant has a two year expected service life under normal use conditions.
The optional motion/nurse call/SmartTV pendant has a two year expected service life under normal use conditions.
The optional
i
Bed
Wireless components have a three year expected service life under normal use conditions.
Contraindications
None known.
Specifications
Safe working load
Note:
Safe working load indicates the sum of the
occupant and accessory weight.
500 lb
227 kg
Product weight
570 lb
259 kg
Scale system capacity (optional equipment), loads weighing up to:
500 lb
227 kg
Scale system accuracy (optional equipment)
± 3 lb for patients weighing 50 to 100 lb
± 3% of the total patient weight for patients
weighing 100 to 500 lb
Overal length and width
Siderails up
95 in. x 41.5 in.
241.3 cm x 105.4 cm
Siderails down
95 in. x 39.5 in.
241.3 cm x 100.3 cm
Patient sleep surface
84 in. x 35 in.
213.4 cm x 88.9 cm
Bed height to top of seat litter with 6” casters
Standard bed
16 in. to 30 in.
40.6 cm to 76.2 cm
With battery backup
/
Zoom®
®
19.5 in. to 30 in.
49.5 cm to 76.2 cm
Litter platform to top of siderail
(full up)
Head end siderail
15 in.
38.1 cm
Foot end siderail
15.5 in.
39.37 cm
Space between siderails (full up)
2.5 in.
5.72 cm
Gatch position
0° to 45°
Fowler position
0° to 60°
Trendelenburg and Reverse Trendelenburg
+10° to -10° ± 1°
Electrical requirements - all electrical requirements meet UL 2601
specifications
115 VAC, 60 Hz, 8 A
Outlet option
110 VAC, 60 Hz, 10 A
Battery voltage
(2) 12V 31Ah - Battery backup /
Zoom
(gel
cell battery)
(1) 9V - Nurse call option (alkaline battery)
Summary of Contents for S3 MEDSURG BED
Page 2: ...sample text ...