6209-001-708 Rev-E
EN
www.stryker.com
15
Guidance and manufacturer's declaration - electromagnetic immunity
The System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) is intended for use in the electromagnetic
environment specified below. The customer or the user of the System 6 Stryker Precision Oscillating Tip Saw (REF
6209-000-000) should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
N/A
N/A
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
N/A
N/A
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
N/A
N/A
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
Power frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE:
U
T
is the alternating current mains voltage prior to application of the test level.