Troubleshooting
WARNING:
DO NOT disassemble or service this equipment.
NOTE:
For service, contact your Stryker sales representative or call Stryker customer service. Outside the US,
contact your nearest Stryker subsidiary.
PROBLEM
CAUSE
ACTION
Handpiece does not run or
oscillates at a reduced speed
making cutting difficult.
Battery pack is discharged.
Recharge the battery pack in Stryker
charger.
Battery pack is expended.
Replace the battery pack.
FAST/STD/Safe control is in the safe
mode position.
Slide the FAST/STD/Safe control to the
FAST or STD mode position.
Cartridge is malfunctioning.
Replace the cartridge.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Motor runs but cartridge does
not move.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Cartridge is malfunctioning.
Replace the cartridge.
Battery pack becomes
unusually hot during use.
Circuitry is malfunctioning.
Check the battery pack on the Stryker
charger and replace the battery pack
if required. See the instructions for use
supplied with the battery charger.
Handpiece has become noisy
and vibrates.
Cartridge is not fully seated.
Reseat the cartridge.
Cartridge is malfunctioning.
Replace the cartridge.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Sporadic electrical interference
is experienced.
Electrical noise is present.
Turn off all electrical equipment not in
use in the operating room.
Relocate electrical equipment; increase
spatial distance.
Plug operating room equipment into
different operating room outlets.