
Page 27
Troubleshooting Tips
For a complete list of troubleshooting tips, please
consult the LIFEPAK 15 Operating Instructions
LIFEPAK 15
Monitor/Defibrillator
Temperature
Observation
Action
Screen remains blank
Temp outside range of 24.8° and
45.2° C (76.6° and 113.4° F)
CHECK SENSOR message
appears and value is (---)
• Temp outside of range, the probe
is dislodged, not connected or the
cable not connected to the monitor
• Check that probe is properly
positioned
CHECK SENSOR message
appears while value is
displayed
Temp probe is dislodged and value
is below 31° C (87.8° F) or above
41.0° C (105.8° F)
TEMP: ACCURACY
OUTSIDE LIMITS message
appears and value is XXX
• Temperature module is not
calibrated or accuracy check failed
• Turn device off then on again. If
problem persists, contact qualified
service personnel.
NOTE:
The temperature probe may require 3 minutes
to equilibrate after placement on the patient
monitoring site.
Summary of Contents for LIFEPAK 15
Page 2: ......