Page 15
LIFEPAK 15
Monitor/Defibrillator
Auxiliary power management
1. Connect power cord to power adapter and auxiliary
power source
2. Confirm LED strip on power adapter is illuminated
3. Connect power adapter output cable to power adapter
4. Connect green end of output cable to auxiliary power
connector on back of defibrillator
5. Confirm
(power connected) and
(battery
charging) LEDs are illuminated
Note:
At least one battery should be installed at all times.
Keep monitor/defibrillator connected to auxiliary power
whenever possible to maintain battery charge level.