Page 11
LIFEPAK 15
Monitor/Defibrillator
SpO
2
1. Press ON
2. Connect the pulse
oximeter cable to the
monitor and sensor
3. Attach sensor to ring
finger of the non-
dominant hand and
orient so cable is on
back of patient’s hand
4. Observe the pulse bar
for fluctuation, amplitude
of the pulse bar indicates
relative signal quality
5. Confirm SpO
2
reading
appears and is stable
SpCO & SpMet
Perform steps 1-5, then:
6. Verify that SpCO / SpMet
sensor is in use; ONLY
Rainbow
®
sensors are
capable of reading
SpCO / SpMet
7. Rotate Speed Dial to
outline SpO
2
area
8. Select parameter
9. Select SpCO or SpMet.
Selected value displays
for 10 seconds.
CO2
SpO2
NIBP
ECG
P1
P2
SpO
2
/ SpCO / SpMet
monitoring procedure
Note:
If SpCO or SpMet reading is above
normal limits (SpCO >10%, SpMet >3%)
an Advisory message occurs.