9.

 

Note: Take care to avoid injury from the blade tip while sliding the torque wrench onto or off of the shaft. 

 
Operation 

 

1.

 

Connect the assembled Hand Piece and instrument to the generator and turn the generator power on. 

Note: MIN is the only mode that can be adjusted. 

2.

 

Select the desired minimum power level using the INCREASE and DECREASE "buttons on the generator touchscreen. 
Note: The recommended minimum starting power level is Level 3 (Illustration 4). For greater tissue cutting speed use a 

higher generator power level, and for greater coagulation use a lower generator power level. The amount of energy 
delivered to the tissue and resultant tissue effects are a function of many factors, including the power level selected, blade 
characteristics, grip force, tissue tension, tissue type, pathology, and surgical technique. 

3.

 

For optimal performance and to avoid tissue sticking, clean the instrument blade, clamp arm, and distal end of the shaft 
throughout the procedure by activating the instrument tip in saline (Illustration 5). 
Note: Do not touch the instrument to metal while activated (Illustration 6).  
Note: Do not clean the blade tip with abrasives. It can be wiped with a moist gauze sponge to remove tissue, if necessary. 

If tissue is still visible in the clamp arm, use hemostats to remove residue, taking care not to actuate the Hand Piece. If 
desired, the instrument may be unplugged (Illustration 7).  

4.

 

The blade is ultrasonically energized when either the foot switch pedal is depressed or one of the hand control buttons is 

depressed. Pressing the left foot pedal of the footswitch or the lower hand control button (MIN) on the instrument 
activates the selected minimum power level. Pressing the right foot pedal of the footswitch or upper hand control button 
(MAX) on the instrument activates the maximum power level. Pressing the Advanced Hemostasis button on the 

instrument activates the Advanced Hemostasis. Under default settings, Advanced Hemostasis is hand-activated only. Note: 
The "MIN" button can be re-assigned to Advanced Hemostasis by selecting "Convert MIN to Advanced Hemostasis" on the 

GEN11 "Settings” screen. Refer to the Generator G11 Operator's Manual "Settings" section (Illustration 8). Once "Convert 

MIN to Advanced Hemostasis" is selected, the Advanced Hemostasis button is disabled. 
Note: Energy is not delivered in Advanced Hemostasis mode until the jaws are completely closed. 

Caution: 

If activation is unintentionally stopped while sealing, maintain jaw closure and reactivate. 

Note: The generator provides an audible tone to indicate when the instrument blade is active. The generator changes to a 

second activation tone as Adaptive Tissue Technology regulates the delivery of energy. 

 

Thermal influences such as fluids or minimal to no tissue in the jaws may affect the presence or timing of the 

tone change. 

 

The tone change does not provide confirmation of tissue effect. When the second tone is heard, the situation 

should be assessed and the intended surgical action completed, such as gradual application of tension to 
facilitate transection. 

 

The secondary activation tone change is not a substitute for surgical experience. 

Note: Scratches on the blade may lead to premature blade failure. 

 

Avoid accidental contact with other instruments during use. 

 

Do not use any means other than the torque wrench to attach or detach the instrument from the Hand Piece. 

5.

 

Close the clamp arm by closing the trigger, and insert the shaft through a trocar or incision. 

6.

 

The instrument can be used for dissection, grasping, coagulation, and cutting between the blade and clamp arm. 
Note: To achieve complete sealing, the trigger should be fully closed and the vessel fully contained between clamp arm and 

blade of device. An audible and tactile "click" indicates full trigger closure. To achieve full closure of the jaws of the device, 
squeeze the plastic trigger until you feel it stop against the plastic handle (plastic to plastic) (Illustration 9). If full trigger 
closure is released prior to or during activation on tissue, an audible and tactile "click" is evident. Increase grip force until 

full trigger closure is achieved. 
Note: Keep clamp arm open when using the inside bottom of the blade for backcutting (Illustration 10).  

WARNING:

 Do not use Advanced Hemostasis mode for procedures where energy application is desired prior to full 

closure of the jaws (e.g. solid organs). Energy is not delivered in Advanced Hemostasis mode until the jaws are completely 

closed. 

WARNING: 

During benchtop testing of vessels >5 mm, the strongest vessel seals were achieved by allowing the Advanced 

Hemostasis mode to completely transect the targeted vessel. 

WARNING:

 Prolonged usage of Advanced Hemostasis Mode may cause tissue pad damage. 

7.

 

Close the clamp arm by closing the trigger and remove the shaft from the trocar or incision. 

 

Shaft Rotation 

1.

 

The instrument’s shaft can be rotated 360°to facilitate visualization and access to target tissue when dissecting, grasping, 

coagulating, and cutting. 

Summary of Contents for Harmonic Ace+7

Page 1: ...se Caution Federal U S A law restricts this device to sale by or on the order of a physician STERILE Explanation of Symbols Sterilized by Ethylene Oxide Gas Date of Reprocessing Use by Date Product Code Do Not Reuse See Instructions For Use Stryker Sustainability Solutions Inc 2016 1810 W Drake Dr Tempe AZ 85283 sustainability stryker com 888 888 3433 ...

Page 2: ...Reprocessed HARMONIC ACE 7 5mm Diameter Shears with Advanced Hemostasis Page 2 of 19 ...

Page 3: ...Reprocessed HARMONIC ACE 7 5mm Diameter Shears with Advanced Hemostasis Page 3 of 19 ...

Page 4: ... represent relative vessel size The MAX button is typically used for smaller vessels where cutting speed is fastest The MIN button is typically used in slightly larger vessels and has reduced cutting speed It is indicated for vessels up to 5 mm in size The Advanced Hemostasis button is designed for larger vessels and is indicated for vessels up to 7 mm in size In this mode cutting speed is further...

Page 5: ...em failure ensure the availability of the appropriate back up equipment relevant to the specific procedure Audible high pitched ringing resonating from the blade or Hand Piece are an abnormal condition and an indicator that the blade or Hand Piece is not operating properly The ringing may be an indicator that the Hand Piece is beyond its useful life or that the blade has not been attached properly...

Page 6: ... the following functions of Adaptive Tissue Technology are no longer available Regulated Energy Delivery Enhanced Audible Feedback and Advanced Hemostasis As a consequence the device vessel sealing indication does not exceed 5 mm Instruments or devices which come into contact with bodily fluids may require special disposal handling to prevent biological contamination Incidental and prolonged activ...

Page 7: ...tasis mode until the jaws are completely closed Caution If activation is unintentionally stopped while sealing maintain jaw closure and reactivate Note The generator provides an audible tone to indicate when the instrument blade is active The generator changes to a second activation tone as Adaptive Tissue Technology regulates the delivery of energy Thermal influences such as fluids or minimal to ...

Page 8: ...ucts Stryker warrants all reprocessed products subject to the exceptions provided herein to be free from defects in reprocessing and to substantially conform to the product specifications contained in the documentation provided by Stryker with the products for one use in accordance with the instructions for use of such product STRYKER SHALL NOT BE LIABLE FOR ANY DAMAGES TO THE EXTENT CAUSED BY ANY...

Page 9: ...ker determines in its reasonable discretion that the claimed defect or non conformance in the product is excluded from warranty coverage as described hereunder it will notify the customer of such determination and will provide an estimate of the cost of repair of the product In such an event any repair would be performed at Stryker s standard rates Products and product components repaired or repla...

Page 10: ...édérale américaine É U ce dispositif ne peut être vendu que par un médecin ou sur ordonnance médicale STÉRILE Explication des symboles Stérilisé par oxyde d éthylène Date du retraitement Date de péremption Code du produit Ne pas réutiliser Voir le mode d emploi Stryker Sustainability Solutions Inc 2016 1810 W Drake Dr Tempe AZ 85283 États Unis sustainability stryker com 888 888 3433 ...

Page 11: ...Cisailles HARMONIC ACE 7 retraitées de 5 mm de diamètre pour hémostase avancée Page 11 de 19 ...

Page 12: ...Cisailles HARMONIC ACE 7 retraitées de 5 mm de diamètre pour hémostase avancée Page 12 de 19 ...

Page 13: ...ent peuvent pivoter sur 360 pour faciliter la visualisation et l accès aux tissus ciblés Les trois tirets affichés sur l instrument représentent une dimension relative du vaisseau Le bouton MAX est généralement utilisé pour les vaisseaux plus petits lorsque la vitesse de la coupe est plus rapide Le bouton MIN est généralement utilisé pour les vaisseaux légèrement plus grands et que la vitesse de c...

Page 14: ...ntion chirrugicale Une compréhension approfondie des principes et techniques impliqués dans les procédures laser électrochirurgicales et ultrasoniques est essentielle pour éviter tout risque d électrocution et de brûlure infligée à la fois au patient et au personnel médical ainsi que pour éviter tout dommage au dispositif ou aux autres instruments médicaux Assurez vous que ni l isolation électriqu...

Page 15: ...ne fois l instrument retiré examinez les tissus pour déceler une mauvaise hémostase Si l hémostase est inadéquate utilisez les techniques appropriées pour obtenir l hémostase Une hémostase fructueuse peut demander des mesures supplémentaires lorsque les instruments HARMONIC sont utilisés sur des organes pleins En raison de la difficulté de visualisation des structures internes procédez lentement e...

Page 16: ...niveau de puissance minimal voulu à l aide des boutons INCREASE augmenter et DECREASE diminuer qui apparaissent à l écran du générateur Remarque Le niveau de puissance minimal de démarrage recommandé est 3 Illustration 4 Pour une vitesse accrue de coupe des tissus utilisez un niveau de puissance plus élevé du générateur et pour une coagulation accrue utilisez plutôt un niveau de puissance plus fai...

Page 17: ...te du déclencheur Remarque Maintenez le bras de serrage ouvert lorsque vous utilisez la partie inférieure interne de la lame pour une coupe inverse Illustration 10 AVERTISSEMENT N utilisez pas le mode d hémostase avancée pour les procédures où une application d énergie est requise avant la fermeture complète des mâchoires p ex dans le cas des organes pleins En mode d hémostase avancée l énergie es...

Page 18: ...AUCUN CAS LA RESPONSABILITÉ DE STRYKER EN LIEN AVEC LA VENTE DU PRODUIT QUE CE SOIT DANS LE CADRE DES THÉORIES DE LA RUPTURE D UN CONTRAT D UN TORT D UNE FAUSSE REPRÉSENTATION D UNE FRAUDE D UNE GARANTIE D UNE NÉGLIGENCE D UNE RESPONSABILITÉ STRICTE OU DE TOUTE AUTRE THÉORIE DU DROIT NE PEUT ÊTRE SUPÉRIEURE AU PRIX D ACHAT À LA VALEUR ACTUELLE DU MARCHÉ OU À LA VALEUR RÉSIDUELLE DES PRODUITS SELON...

Page 19: ...mplacé la propriété de Stryker Si un remboursement est fourni par Stryker le produit pour lequel le remboursement est décidé doit être retourné à Stryker et deviendra la propriété de Stryker HARMONIC et HARMONIC ACE sont des marques de commerce d Ethicon Endo Surgery ULS EL10053 Rév E 10 2017 RM702506 Les informations concernant le fabricant d équipement d origine fournies sur l étiquette apparais...

Reviews: