1-52
28EX-009-002 REV B
En
gli
sh
POWER CORD CONNECTOR FUSE REPLACEMENT
Tools Required:
•
Small Regular Screwdriver
Procedure:
Note
Refer to Figure 12,
, for this procedure .
1 . Remove the power cord from the power cord connector .
2 . Using a small regular screwdriver, slide the fuse holder door open .
3 . Remove the defective fuse and replace it with a new one .
Note
120 Vac rated FL28EX beds use fast acting, 250 Vac, 10A type fuses . For beds rated at voltages other than 120 Vac,
see “Specifications”,
, for fuse ratings .
4 . Close the fuse holder door .
Note
Make sure to use the right fuse holder for the fuses used with the bed and make sure the fuses are correctly installed
to be stable in the fuse holder .
5 . Test the bed for proper operation before returning the bed to service .
BED EXIT BEEPER REPLACEMENT
Tools Required:
•
#2 Phillips Screwdriver
Procedure:
Note
Refer to Figure 13,
, for this procedure .
1 . Raise the bed to the highest position and apply the brakes .
2 . Remove the plastic litter cover (if equipped) and raise the fowler to the highest position .
3 . Unplug the power cord from the wall outlet .
4 . Using a #2 Phillips screwdriver, remove the two screws (A) holding the cover plate .
5 . Properly ground yourself (see “Grounding Diagram,
) .
6 . Disconnect the beeper cable from J3 on the scale board (B) .
7 . Lift the beeper (C) to remove it from the plate . Remove any glue residue .
8 . Reverse the above steps to install the new beeper .
9 . Test the Bed Exit system to ensure the beeper for proper operation before returning the bed to service .
Service Information
Summary of Contents for FL28EX
Page 2: ......
Page 83: ...www stryker com 28EX 009 002 REV B 1 81 Return To Table of Contents English Notes ...
Page 99: ...www stryker com 28EX 009 002 REV B 1 97 Return To Table of Contents English Notes ...
Page 162: ......
Page 257: ...www stryker com 28EX 009 002 REV B 2 95 Retour à la table des matières Français Commentaires ...
Page 320: ......
Page 321: ......