1-30
28EX-009-002 REV B
En
gli
sh
HEAD END SIDERAIL ASSEMBLY REPLACEMENT
Tools Required:
•
1/2” Socket
•
3/8” Drive Ratchet
•
3/16” Allen Wrench
•
Cutting Pliers
Procedure:
1 . Raise the bed to the highest position and apply the brakes .
2 . If applicable, remove the plastic litter cover (if equipped) . Raise the fowler to the highest position and raise the
siderail to be repaired .
3 . Unplug the power cord from the wall outlet .
4 . Loosen the lock ring (A) and unplug the siderail cable .
5 . Using cutting pliers, remove the cable ties holding the siderail cable to the frame .
6 . Using a 3/16” Allen wrench and a 1/2” socket, remove the four locknuts/bolts (B) holding the siderail assembly to
the head section and remove the assembly . Support the assembly when removing the last bolts .
7 . Reverse the above steps to install the new siderail assembly .
8 . Test the siderail movement and all controls of both control panels, including the nurse call (optional) and the
communications package (optional), for proper operation before returning the bed to service .
A
B
B
Figure 1