19
Weigh System Usage
ADDING OR REMOVING ITEMS DURING A PATIENT’S STAY
If it is necessary to add or remove items (monitors, pumps, etc.) during the patient’s stay, press and re-
lease
to activate the weigh system. After the scale monitor reads: “XXX.X LB”, press and hold
The scale monitor will read:
“HOLD TO START”
“RELEASE TO START”
Release
The scale monitor will read:
“DO NOT TOUCH BED”
“ADD/REMOVE EQUIP.”
Add or remove the equipment and press
The scale monitor will read:
“RELEASE TO FIN.”
Release
The scale monitor will read:
“DO NOT TOUCH BED”
“XXX.X LB”
The weight displayed will be that of the patient only.
If the CHANGE EQUIPMENT function is started but not finished, after 50 seconds the monitor will read:
“HIT CH. EQ. TO END”
Press
The scale monitor will read:
“RELEASE TO FIN.”
Release
The scale monitor will read:
“DO NOT TOUCH BED”
“XXX.X LB”
Summary of Contents for Epic