![Stryker Epic Operation Manuals Download Page 18](http://html.mh-extra.com/html/stryker/epic/epic_operation-manuals_1385889018.webp)
17
Foot Board Operation Guide
WEIGH SYSTEM CONTROL PANEL GUIDE
1.
LCD – displays patient weight. Trendelenburg angle is displayed when the scale is not active.
2.
Press to zero bed (see page 18). Also press to scroll while Menu Mode is active.
3.
Press to enter and exit the Menu Mode.
4.
Press when adding or removing equipment to the bed (see page 19).
5.
Press to turn weigh system on and off. Also press to scroll while Menu Mode is active.
6.
Press to change weight from pounds to kilograms or back (see page 20). Also press while using the Menu
Mode.
7.
Press to display the Trendelenburg or Fowler angle of the bed (see page 20).
NOTE
If weight is displayed, SCALE ON/OFF must be pressed to turn off the scale before the Trend. or Fowler angle
will display.
WARNING
Scale function may be affected by siderail/caster interference. With the litter fully lowered or lowered in Re-
verse Trendelenburg, the siderails tucked under the litter in the storage position and the casters turned, there
is the potential for interference between the siderail and the caster. Raise the siderails when lowering the
litter to the full down position to prevent the interference from causing the bed’s scale system to weigh inaccu-
rately.
For more detailed operating instructions see the following:
1.
Preparing The Bed For Patient Stay – page 18
2.
Activating the Weigh System and Displaying Patient Weight – page 18
3.
Adding or Removing Items During a Patient’s Stay – page 19
4.
Displaying Trendelenburg or Fowler Angle – page 20
5.
Converting the Patient’s Weight – page 20
6.
Viewing Patient Weight In Gain/Loss Mode – page 21
7.
Changing the Numerical Value Of Displayed Weight – page 22
This panel is optional equipment.
Summary of Contents for Epic
Page 24: ...Optional Pendant Operation 23 ...