Stryker Epic Operation Manuals Download Page 16

15

Foot Board Operation Guide

FOOT BOARD CONTROL PANEL GUIDE (CONTINUED)

1.

Press to raise back section.

2.

Press to raise knee section.

3.

Press to lower back section.

4.

Press to lower knee section.

1. Press to activate the emergency CPR drop function.  The bed will level from Trendelenburg/reverse Tren-

delenburg, the Fowler will lower to flat, the Knee will lower to flat and the litter will lower to full down.

2. Press to activate the Cardiac Chair function.  The Knee will raise, the Fowler will raise or lower to approxi-

mately 52

 and the bed will tilt to approximately –12

 reverse Trendelenburg (foot end down) or –14

if the bed has the enhanced height option.  Release the button to stop bed movement: hold the button until
movement stops to complete the function.

Summary of Contents for Epic

Page 1: ...Critical Care Bed Model 2030 2031 OPERATIONS MANUAL For Parts or Technical Assistance 1 800 327 0770 ...

Page 2: ... Preparing the Bed for Patient Stay 18 Adding or Removing Items During a Patient s Stay 19 Displaying Trendelenburg or Fowler Angle 20 Converting the Patient s Weight 20 Viewing the Patient s Weight in Gain Loss Mode 21 Changing the Numerical Value of Displayed Weight 22 Optional Pendant Operation 23 Optional Epic Head End Control Panel Operation 24 Battery Charging and Operation Epic Option 25 Op...

Page 3: ...casters 19 9 to 34 5 50 5 cm to 88 cm 6 casters 21 9 to 36 5 56 cm to 93 cm 8 casters Knee Gatch Angle 0 to 35 Back Angle 0 to 90 Trendelenburg Reverse Trendelenburg 14 to 14 Electrical Requirements 115 VAC 60 Hz 7 0 Amps Battery Voltage Optional 24 V 31 Ah Noise Level 65 Decibels Stryker reserves the right to change specifications without notice WARNING CAUTION NOTE DEFINITION The words WARNING C...

Page 4: ...udes an additional safety precaution in that means are provided for the connection of the EQUIPMENT to the protective earth con ductor in the fixed wiring of the installation in such a way that ACCESSIBLE METAL PARTS cannot become live in the event of a failure of the BASIC INSULATION IPX4 Protection from liquid splash Dangerous Voltage Symbol Protective Earth Terminal Potential Equalization Symbo...

Page 5: ... push on the bed to ensure the brakes are locked Injury could result if the bed moves while a patient is getting in or out of bed Ensure the brakes are completely released prior to attempting to move the bed Attempting to move the bed with the brakes actuated could result in injury to the user and or patient Assistance is required to lower the Back if the angle of the Back is greater than 80 when ...

Page 6: ...n to safety features including but not limited to Safety side latching mechanisms Caster braking systems Leakage current 100 microamps max No controls or cabling entangled in bed mechanisms Frayed electrical cords and components All controls return to off or neutral position when released Always unplug bed during service or cleaning When working under the bed always place blocks under the litter f...

Page 7: ...e siderails raise and lower smoothly and lock in the up and intermediate positions Run through each function on the foot board control panel and ensure that each is working properly see function lockout system usage page 14 and weigh system control panel guide page 17 Ensure all functions are working properly on the siderail controls Raise the Back up to approximately 60 Squeeze the CPR release ha...

Page 8: ...derails that will blink when the brakes are not engaged only if the bed is plugged into a wall socket see page 10 The brakes will still operate properly when the bed is not plugged in If the bed is equipped with the Epic option there is also a Release Brakes LED on the Head End Control Panel that will illuminate when the brakes are engaged while the battery power switch is on see page 24 STEER PED...

Page 9: ...d can also be removed and used as an emergency CPR board FOLEY BAG HOOKS USAGE The standard Foley bag hooks are found at two locations on both sides of the bed under the frame rail below the seat section and at the extreme foot end of the bed The patient weight reading on the bed scale system will not be affected when the Foley bag hooks are used FOOT PROP USAGE To prop the foot end of the Knee up...

Page 10: ...verse Trendelenburg the siderails tucked under the litter in the storage position and the casters turned there is the potential for interference between the siderail and the caster Raise the siderails when lowering the litter to the full down position to prevent the interference from causing the bed s scale system to weigh inaccu rately CAUTION The siderails are not intended to be used as a push d...

Page 11: ...nches 9 Press to lower the litter 10 There are two prevention mattress systems available with two different sets of siderail controls If the bed is equipped with a Dynamic Mattress System press to activate the automatic or manual operation of the DMS The LED will light to indicate which mode is activated 11 If the bed is equipped with a StryKair Mattress press to lock out patient control of the St...

Page 12: ... to increase the TV or radio volume 8 Press to decrease the TV or radio volume 9 Press to turn on the room lights Press again to turn off 10 Press to turn on the reading light Press again to turn off 11 Press to increase the firmness of the mattress 12 Press to decrease the firmness of the mattress WARNING When attaching equipment to the bed ensure it will not impede normal bed operation For examp...

Page 13: ... lights Press again to turn off 10 Press to turn on the reading light Press again to turn off 11 Press to change the TV channel up 12 Press to change the TV channel down 13 Press to mute the TV sound Press again to turn the sound back on 14 Press to display TV closed captioning 15 Press to increase the firmness of the mattress 16 Press to decrease the firmness of the mattress WARNING When attachin...

Page 14: ... Press again to unlock 5 Press to lock out bed up down motion controls on the siderails Press again to unlock 6 Press to raise the bed height If your bed is equipped with the enhanced height option continue to hold the button an additional 5 seconds after the first stop The litter will raise an additional 2 inches 7 Press to lower bed 8 Press to lower head end of bed Trendelenburg 9 Press to lower...

Page 15: ...tion switches on the bed press the ON OFF switch in the Siderail Control Lock outs module The padlock symbol on the control panel will be lighted when that function is locked out 2 To lock out the bed movement functions on the siderails and prevent the patient from changing the posi tioning of the bed press the BACK or KNEE switch in the Siderail Control Lockouts module The padlock symbol on the c...

Page 16: ...lenburg reverse Tren delenburg the Fowler will lower to flat the Knee will lower to flat and the litter will lower to full down 2 Press to activate the Cardiac Chair function The Knee will raise the Fowler will raise or lower to approxi mately 52 and the bed will tilt to approximately 12 reverse Trendelenburg foot end down or 14 if the bed has the enhanced height option Release the button to stop ...

Page 17: ... adding linens and equipment to the bed Press and hold the ARM and DISARM keys together for 5 seconds The ARMED light will begin to flash Release the ARM and DISARM keys and do not touch the bed until the ARMED light stops flashing After the ARMED light stops flashing place the new patient is on the bed Push and release the ARM key The BED EXIT ON light will come on To deactivate Bed Exit push DIS...

Page 18: ...splay WARNING Scale function may be affected by siderail caster interference With the litter fully lowered or lowered in Re verse Trendelenburg the siderails tucked under the litter in the storage position and the casters turned there is the potential for interference between the siderail and the caster Raise the siderails when lowering the litter to the full down position to prevent the interfere...

Page 19: ... TO ZERO WT RELEASE TO ZERO Release The scale monitor will now read DO NOT TOUCH BED 0 0 LB The bed is now ready for the patient NOTE If Bed Exit is armed it must be disarmed before the scales can be zeroed NOTE Do not zero the bed while a patient is in bed If this should occur remove the patient and zero the bed ACTIVATING THE WEIGH SYSTEM AND DISPLAYING PATIENT WEIGHT Press and release The scale...

Page 20: ...ASE TO START Release The scale monitor will read DO NOT TOUCH BED ADD REMOVE EQUIP Add or remove the equipment and press The scale monitor will read RELEASE TO FIN Release The scale monitor will read DO NOT TOUCH BED XXX X LB The weight displayed will be that of the patient only If the CHANGE EQUIPMENT function is started but not finished after 50 seconds the monitor will read HIT CH EQ TO END Pre...

Page 21: ...LE X or FOWLER ANGLE X If scale system is not active press and hold The scale monitor will read LET GO FOR FOWL FOWLER ANGLE X or LET GO FOR TREND TREND ANGLE X CONVERTING THE PATIENT S WEIGHT Press and release The scale monitor will read WEIGHT NOW KGS XXX X KG Repeat the procedure to return to pounds The display will read WEIGHT NOW LBS XXX X LB ...

Page 22: ...CALE Press to enter the Menu Mode The scale monitor will read FOR OPTIONS Press to scroll menu options The scale monitor will read START GAIN LOSS PUSH ENTER Press The scale monitor will read DO NOT TOUCH BED XXX X LB G XX X To exit Gain Loss mode press to enter the Menu Mode The scale monitor will read FOR OPTIONS Press to scroll menu options The scale monitor will read QUIT GAIN LOSS PUSH ENTER ...

Page 23: ...ead LET GO FOR SCALE XXX X LB Press to enter the Menu Mode The scale monitor will read FOR OPTIONS Press to scroll menu options The scale monitor will read CHNG PTNT WGT PUSH ENTER Press The scale monitor will read HOLD TO INC TO DEC Press to decrease the displayed weight or to increase the displayed weight Once the desired weight is displayed press The scale monitor will read XXX X LB ...

Page 24: ...Optional Pendant Operation 23 ...

Page 25: ...n additional 5 seconds after the first stop The litter will raise an additional 2 inches 2 Press and hold to lower the litter 3 Press to raise the Knee section 4 Press to lower the Knee section 5 Press to raise the Back section 6 Press to lower the Back section 7 The Plug Bed In To Charge LED will be illuminated while the battery power switch is on if the battery level is low Plug the bed power co...

Page 26: ...tially discharged but still ade quate for operation and only one will flash if the batteries are near full discharge The Plug Bed In To Charge LED on the Head End Control Panel see page 24 will remain illuminated while the battery power switch is on if the battery level is low Plug the power cord into a wall socket to charge the batteries 3 After one hour on battery power the unit will enter power...

Page 27: ... it locks into place at its fully raised position 3 Rotate the I V hangers B to desired position and hang I V bags 4 To lower the I V pole turn the latch C clockwise until section A lowers CAUTION The weight of the I V bags should not exceed 40 pounds To use the Removable I V pole 1 Install the pole at any of the four receptacles on the bed top located on all four corners of the bed 2 To raise the...

Page 28: ...wer light located on the foot board is flashing the Nurse Call battery needs to be replaced The battery is located on the patient s left side at the head end of the bed No tools are required to replace the battery Unplug the bed power cord from the wall socket and replace the battery MAIN BED POWER CIRCUIT BREAKER In the event of a loss of bed functions unplug the bed power cord from the wall sock...

Page 29: ... V pole working properly Foley bag hooks intact Chart rack intact and working properly CPR board not cracked or damaged and stores properly No cracks or splits in head and foot boards All functions on footboard working properly including LED s No rips or cracks in mattress cover Scale and Bed Exit system calibrated properly optional equipment Motion interrupt switches working properly Power cord n...

Page 30: ...es 1 2 ounce gallon Matar Phenolic Huntington Laboratories 1 2 ounce gallon Omega Quaternary Airwick Professional Products Division 1 2 ounce gallon Quanto Quaternary Huntington Laboratories 1 ounce gallon Sanikleen Quaternary West Chemical Products 2 ounces gallon Sanimaster II Quaternary Service Master 1 ounce gallon Vesphene Phenolic Vestal Laboratories 1 1 4 ounce gallon Quaternary Germicidal ...

Page 31: ...R INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING FROM OR IN ANY MANNER RELATED TO SALES OR USE OF ANY SUCH EQUIPMENT To Obtain Parts and Service Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives These representatives are factory trained available locally and carry a substantial spare parts inventory to minimize repair time Simply call your local...

Page 32: ...N THE DELIVERY RECEIPT AT THE TIME OF RECEIPT Upon prompt notification Stryker will file a freight claim with the appropriate carrier for damages incurred Claim will be limited in amount to the actual replacement cost In the event that this information is not received by Stryker within the fifteen 15 day period following the delivery of the merchandise or the damage was not noted on the delivery r...

Page 33: ... 100 4 REV F European Representative Stryker France Phone 33148632290 BP 50040 95946 Roissy Ch de Gaulle Fax 33148632175 Cedex France 6300 Sprinkle Road Kalamazoo MI 49001 9799 USA 800 327 0770 ww strykermedical com ...

Reviews: