33
Preventative Maintenance
MAINTENANCE CHECKLIST
All fasteners secure
Engage brake pedal and push on the bed to ensure all casters lock securely
Optional locking steer caster engages and disengages properly
Siderails move, latch and stow properly
All functions on siderails working properly (including LED’s)
Head End Control Panel working properly (including LED)
−
optional equipment
Confirm battery powered functionality
−
optional equipment
CPR release working properly
Foot prop intact and working properly
I.V. pole working properly
Foley bag hooks intact
Chart rack intact and working properly
CPR board not cracked or damaged and stores properly
No cracks or splits in head and foot boards
All functions on footboard working properly (including LED’s)
No rips or cracks in mattress cover
Scale and Bed Exit system calibrated properly
−
optional equipment
Motion interrupt switches working properly
Power cord not frayed
No cables worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 100 milliohms
Current leakage not more than 300 microamps
Apply grease to litter grease points
Bed Serial No.
Completed By:_________________________________ Date:_____________
Summary of Contents for Epic II 2030
Page 29: ...Optional Pendant Operation 28 ...