5
1. Plug the CastVac power cord into a hospital grade wall
outlet. Plug the other end into the appliance inlet.
NOTES:
• CastVac REF 996 only, plug the source cord into a
standard wall outlet.
• The function switch, located on the front of the canister,
activates the equipment. It is not the power supply switch.
To disconnect the vacuum from the power supply, unplug it
at the wall receptacle.
• CastVac REF 996 can be unplugged at the vacuum's
appliance inlet located at lower back portion of the
canister.
2. Plug the Stryker Cast Cutter into the receptacle located
on the side of the canister. See illustration. Wrap excess
cord length around the handles of the CastVac Stand.
CAUTION:
This receptacle is designed to accept Stryker
Cast Cutter only. DO NOT attach other devices.
The switch in the "ON" position enables the CastVac™ to run
constantly.
The Cast Cutter operates independently and can be turned on and off
while the vacuum runs steadily.
Models REF
986, 987 & 996
With the switch in the "STANDBY" position, both the Cast Cutter and
CastVac are activated by the Cast Cutter switch.
The vacuum runs only when the Cast Cutter is switched on and shuts
down when the Cast Cutter is switched off.
NOTE:
CastVac model REF 996 will continue to run 10 seconds after
the Cast Cutter is switched "OFF".
With the switch in the "OFF" position, only the Cast Cutter will run when
activated.
The CastVac remains off while the Cast Cutter is running.
The 3-way switch can be set for the following modes of operation:
Operation
Model
941
only.
Model 996
3. Operate the system in the desired mode by setting the 3-way switch located on the lower front portion of the canister.
(See special instructions below).