6
WARNING:
To reduce the risk of electric shock or injury, unplug the CastVac prior to
any maintenance activity.
1. Release the two latches located at the sides of the vacuum canister. Remove canister lid.
2. Position the High Performance Filter Cartridge in the approximate center of the cannister.
The filter opening must face up.
3. Place the lid over the canister so that the hose port in the lid and filter interface.
4. Press the lid down and twist hose port to seat in High Performance Filter Cartridge. Fasten
latches.
5. Dispose of used filter in a suitable manner.
CAUTIONS:
• Allow vacuum to dry completely before using.
• DO NOT IMMERSE any portion of the vacuum unit.
1. Unplug vacuum unit before cleaning.
2. Use a sponge dampened in soapy water or disinfectant to clean all external areas of the canister and stand. DO NOT allow
liquids to splash or seep inside the unit.
No adjustments are necessary. However, it is suggested that prior to use, the vacuum unit should be operated and inspected for any
damage. DO NOT use if damage is apparent.
Filter Replacement
Cleaning & Maintenance
Periodic Maintenance
Activity/Item
Source cord
Interval
Annually
Tools and Equipment Required
N/A
Instructions Models REF 986 & 987
Check for cracks. Return to Stryker
for for repair if required.
Instructions Model REF 996
Check for cracks. Replace
cord.