6
www.stryker.com
EN
7110-120-700 Rev-AB
Instructions
To Connect the Battery Charger
WARNINGS:
• To avoid the risk of electric shock, ALWAYS
connect this equipment to a hospital-grade,
facility power receptacle with protective earth.
• DO NOT modify the power cord.
• ALWAYS position the equipment so that the
power cord may be easily disconnected as
required.
• DO NOT place the battery charger within the
sterile field.
• DO NOT touch the battery charger and the
patient simultaneously. Failure to comply may
cause electrical shock.
CAUTION:
ALWAYS place the power cord where it
will not be stepped on, tripped over, or otherwise
subjected to damage or stress.
1. Use the power cord to connect the battery
charger to a hospital-grade, facility power
receptacle (figure 1). The power-up sequence
will begin:
-
A diagnostic test will be performed to
ensure the battery charger and each
installed charger module is operational
-
The battery charger software and hardware
revision numbers will appear (figures 2
and 3)
-
The charger module revision number will
appear for each installed charger module
(figure 4).
NOTE:
If an ERROR message appears during
the power-up sequence, note the error code that
appears in the counter area (figure 5). See the
Troubleshooting
section.
2. If desired, use an Ethernet cable (not
supplied) to connect the isolated diagnostics
port to a facility Ethernet receptacle.
NOTES:
• Confirm with a facility IT representative
that the Ethernet receptacle has access to
Stryker’s Cloud.
• See the
Network Requirements
in the
Specifications
section.
• For more information, contact your Stryker
sales representative.
Figure 1 – Battery Charger Connections
Figure 2 – Software Revision Number
Figure 3 – Hardware Revision Number
Figure 4 – Charger Module Revision Number
Figure 5 – Error Code