12
www.stryker.com
EN
7110-120-700 Rev-AB
PROBLEM
COMPONENT
CAUSE
ACTION
NO BATTERY appears in
the message area of the
information screen when a
battery is installed.
Battery pack
Battery pack is not
properly seated in
charger module.
Remove and reinstall the
battery pack.
Stryker F1 battery
pack release button
is not allowing proper
installation.
Push the release button
during installation.
Stryker F1 battery pack
is not installed in the
correct orientation.
Reinstall battery pack in
the correct orientation.
Contacts are dirty or
corroded.
Clean the battery pack.
Battery pack is dead.
Replace the battery
pack.
Charger module
Charger module is
damaged.
Replace the charger
module.
Message area of the
information screen cycles
between NO BATTERY and
CHARGING.
Battery pack
Battery pack has low
voltage.
Leave the battery pack
on the charger module
for a minimum of five
minutes. If the problem
persists, check other
causes.
Battery pack is dead.
Replace the battery
pack.
REPLACE appears in
the message area of the
information screen.
Battery pack
Battery pack is wet.
Dry and reinstall the
battery pack.
Battery pack has
exceeded its operational
life.
Replace the battery
pack.
Charger module
Charger module contact
is wet.
Dry the contact and
reinstall the battery pack.
Internal problem.
Reinsert the battery. If
the REPLACE message
does not occur again,
then replace the charger
module.
Battery pack does not fit
into charger module.
Charger module
Charger module is
intended for different
battery pack.
Install the proper
charger module. See
the instructions for use
supplied with the charger
module.
Charger module is loose.
Charger module
Screw is not secure.
Tighten the screw.