7110-120-700 Rev-AB
EN
www.stryker.com
13
PROBLEM
COMPONENT
CAUSE
ACTION
Battery pack becomes
unusually hot during use or
while charging.
Charger module
Battery pack is intended
for different charger
module.
Install the battery pack
in the correct charger
module.
Battery pack
Internal problem.
Check battery pack
status in the battery
charger; replace the
battery pack if indicated.
Battery charger
Internal problem.
Return the battery
charger for repair.
Information screen does not
display battery pack state-
of-health information.
Battery charger
The battery charger will
display the battery pack
state-of-health at the end
of the charge cycle.
Wait until the charge
cycle is complete.
Information screen does not
display charge cycle count
information.
Battery pack
Battery pack is not a
Stryker-approved battery
pack.
Use Stryker battery
packs only.
Battery pack/
Charger module
Battery pack (REF 4112-
000-000) is not a design
integrated component.
No action required.
Amber indicator light
flashes continuously.
Battery pack/
charger module
Battery pack and/or
battery charger contacts
may be dirty.
Clean the contacts.
ERROR code 005 appears
in the counter area of the
information screen.
Battery pack
Battery pack is wet.
Dry and reinstall the
battery pack.
Charger module
Charger module contact
is wet.
Dry the contact and
reinstall the battery pack.
Charger module contact
is corroded.
Clean the charger
module contact. If
problem persists, replace
the charger module.