Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be no closer than 12 inches (30 cm) to any part of the S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt locator, including cables specified by the
manufacturer.
•
Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the product. If such
use is necessary, carefully observe stacked or adjacent equipment to make sure that they operate properly.
•
The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
The 521200380100 S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt locator was evaluated using the following cables:
C
Ca
ab
blle
e
L
Le
en
ng
gtth
h ((m
m))
AC mains input cable
1.2
Nurse call (DB-37)
2.4
G
Gu
uiid
da
an
ncce
e a
an
nd
d m
ma
an
nu
uffa
accttu
urre
err’’ss d
de
ecclla
arra
attiio
on
n -- e
elle
eccttrro
om
ma
ag
gn
ne
ettiicc e
em
miissssiio
on
nss
The 521200380100 S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt locator is intended for use in the electromagnetic environment specified below.
The customer or the user of the 521200380100 S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt locator should assure that it is used in such an
environment.
E
Em
miissssiio
on
nss tte
esstt
C
Co
om
mp
plliia
an
ncce
e
E
Elle
eccttrro
om
ma
ag
gn
ne
ettiicc e
en
nvviirro
on
nm
me
en
ntt
RF Emissions
CISPR 11
Group 1
N
No
otte
e -- The emissions characteristics of this
equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B
is normally required) this equipment might not offer
adequate protection to radio-frequency
communication services. The user might need to take
mitigation measures, such as relocating or re-
orienting the equipment.
RF Emissions
CISPR 11
Class A
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Complies
G
Gu
uiid
da
an
ncce
e a
an
nd
d m
ma
an
nu
uffa
accttu
urre
err’’ss d
de
ecclla
arra
attiio
on
n -- e
elle
eccttrro
om
ma
ag
gn
ne
ettiicc iim
mm
mu
un
niittyy
The 521200380100 S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt locator is suitable for use in a professional healthcare facility environment and not
in environments exceeding immunity test conditions that the product was evaluated to, such as near high frequency (HF)
surgical equipment and inside of the radio frequency (RF) shielded room of magnetic resonance imaging (MRI)
equipment. The customer or the user of the 521200380100 S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt locator should assure that it is used in such
an environment and that the electromagnetic environment guidance listed below is followed.
IIm
mm
mu
un
niittyy tte
esstt
IIE
EC
C 6
60
06
60
01
1 tte
esstt lle
evve
ell
C
Co
om
mp
plliia
an
ncce
e lle
evve
ell
E
Elle
eccttrro
om
ma
ag
gn
ne
ettiicc
e
en
nvviirro
on
nm
me
en
ntt--g
gu
uiid
da
an
ncce
e
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrostatic fast transient/
burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Main power quality should
be that of a typical
commercial or hospital
environment.
EN
18
5212-009-205 Rev AB.1