23
Service Information
DISPLAY BOARD REPLACEMENT (CONTINUED)
4. Using two 7/16” wrenches, remove the nuts, bolts and washers (items A, C & F, page 66) holding the display
assembly to the frame.
5. Using a 3/32” hex Allen wrench, remove the 6 set screws under the display weldment. Slide the display
board out toward the (patient) left.
6. Reverse the above steps to install the new display board. Thoroughly test all Modular Warmer operations
before returning the stretcher to service.
FILTER REPLACEMENT
The Thermaltek
t
warming system contains an inlet filter. It is recommended this filter be changed
every four to six months to ensure proper operation of the warming system. See the two blower box
diagrams below for filter part number and location.
Required Tools:
5/32” Hex Allen Wrench, 70825 Torx or Standard Screwdriver
FORMER STYLE
CURRENT STYLE
NOTE
Wire mesh on filter must face internally when installed.
CAUTION
To avoid electrical shock, unplug warmer or stretcher from wall socket before opening blower box.
Summary of Contents for Thermaltek 1030
Page 8: ...Notes 5 2 ...
Page 15: ...Notes 10 2 ...
Page 29: ...24 715 1 250 Side Control Base Assembly with Brakes ...
Page 39: ...Notes 32 2 ...
Page 41: ...Side Control Base Assembly with Jacks 34 Assembly part number 715 1 260 ...
Page 45: ...Side Control Base Assembly With Jacks 37 1 Assembly part number 1210 2 10 ...
Page 48: ...Notes 37 4 ...
Page 49: ...End Control Base Assembly with Jacks 38 FOOT END HEAD END Assembly part number 716 1 270 ...
Page 56: ...43 Notes ...
Page 74: ...57 4 Notes ...
Page 91: ...74 Notes ...
Page 116: ...DH 2 97 1030 90 2 REV A 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...