Service Information
7
SIDERAIL LATCH ADJUSTMENT
Required Tools:
1/8 Hex Allen Wrench
WARNING
The siderail latch adjustment is pre–set at the factory, and there should not normally be a need for readjust-
ment. If adjustment must be done it is important to follow the procedure below. If it is not done properly, injury
to the patient or user could occur.
Adjustment Procedure:
1. Using a 1/8 hex Allen wrench, adjust the hex Allen screw located on the latch assembly opposite the latch.
Turning the Allen screw clockwise will
DECREASE
the amount of ”play” in the latching mechanism. Turn-
ing counterclockwise will
INCREASE
the amount.
NOTE
The amount of ”play” in the siderail, when in full up engaged position, should be approximately 1/8 to 3/16
inches.
CAUTION
Too much ”play” when the siderail is in the full up engaged position will give the siderail the appearance of
being unstable and could also cause premature wearing of the latch system.
Too little ”play” will obstruct the latch and keep it from engaging completely in the full up position, which may
cause damage to the latch and/or injury to the patient or user.
Summary of Contents for Thermaltek 1030
Page 8: ...Notes 5 2 ...
Page 15: ...Notes 10 2 ...
Page 29: ...24 715 1 250 Side Control Base Assembly with Brakes ...
Page 39: ...Notes 32 2 ...
Page 41: ...Side Control Base Assembly with Jacks 34 Assembly part number 715 1 260 ...
Page 45: ...Side Control Base Assembly With Jacks 37 1 Assembly part number 1210 2 10 ...
Page 48: ...Notes 37 4 ...
Page 49: ...End Control Base Assembly with Jacks 38 FOOT END HEAD END Assembly part number 716 1 270 ...
Page 56: ...43 Notes ...
Page 74: ...57 4 Notes ...
Page 91: ...74 Notes ...
Page 116: ...DH 2 97 1030 90 2 REV A 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...