Service Information
12
HYDRAULIC TROUBLESHOOTING
Be sure the pedal linkage has been adjusted properly before beginning service on the jacks (see page 8 or
page 9).
PROBLEM/SYMPTOM
SOLUTION
Jack will not raise to full height.
Add hydraulic fluid (see p.14). Check for leaks.
Jack will not hold in raised position.
Close the needle valve completely. If the jack
holds, replace valve #1 (see p. 15). If the jack
does not hold, replace valve #2 (see p. 16).
Jack will not pump up and the jack actuator rod
does not
move.
Close the needle valve. If the jack will now pump
up, replace valve #1. If the jack still will not pump
up after closing the needle valve, replace valve #3
(see p. 16).
Jack will not pump up but the jack actuator rod
does
move when the pump pedal is activated.
Replace valve #2 (see p. 16).
Jack will not pump up and the jack actuator rod
may or may not move.
Remove excess air (vacuum) in system (see p.
16).
Contact Stryker technical service at 1–800–327–0770 for further assistance.
Summary of Contents for Thermaltek 1030
Page 8: ...Notes 5 2 ...
Page 15: ...Notes 10 2 ...
Page 29: ...24 715 1 250 Side Control Base Assembly with Brakes ...
Page 39: ...Notes 32 2 ...
Page 41: ...Side Control Base Assembly with Jacks 34 Assembly part number 715 1 260 ...
Page 45: ...Side Control Base Assembly With Jacks 37 1 Assembly part number 1210 2 10 ...
Page 48: ...Notes 37 4 ...
Page 49: ...End Control Base Assembly with Jacks 38 FOOT END HEAD END Assembly part number 716 1 270 ...
Page 56: ...43 Notes ...
Page 74: ...57 4 Notes ...
Page 91: ...74 Notes ...
Page 116: ...DH 2 97 1030 90 2 REV A 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...