Service Information
MAIN FUSE REPLACEMENT –
8A Fuse part number 59–57
WARNING
Before servicing the Modular Warmer, unplug the power cord from the wall receptacle.
CAUTION
Before servicing the Modular Warmer, properly ground yourself. (See page 20 for static precautions).
Required Tools:
Standard Screwdriver
Replacement Procedure:
NOTE
Refer to drawing 1030–433–3, Blower Box Assembly, (page 68) for part reference.
1. Using a standard screwdriver, turn the fuse holder (item W, page 68) out of the locked position and remove
it with the attached 8A fuse (item AW, page 68).
2. Remove the old fuse, install the new one and reinstall the fuse holder. Thoroughly test all Modular Warmer
operations before returning the stretcher to service.
DISPLAY BOARD REPLACEMENT – Display Board p/n 1030–80–900
WARNING
Before servicing the Modular Warmer, unplug the power cord from the wall receptacle.
CAUTION
Before servicing the Modular Warmer, properly ground yourself. (See page 20 for static precautions).
Required Tools:
Plastic or Rubber Hammer
3/32” Hex Allen Wrench
(2) 7/16” Comb. Wrenches
Medium Standard Screwdriver
Long Nose Pliers
Replacement Procedure:
NOTE
Refer to assembly drawing 1030–473–10, Display Assembly (page 71) and 1030–426–10, Modular Warmer
Assembly, (page 66) for part reference.
WARNING
The logic control board in this display has been calibrated at the factory according to the test records kept
on file for this warming stretcher. Do not take the calibration shunt out of the old board and place it in the new
one. Do not use this board in any other unit or attempt to alter the temperature settings. Do not use any
existing logic control board in this unit unless it has been factory calibrated for this serial number warming
stretcher. If future service is required, a new board must be ordered with the proper factory testing and calibra-
tion. Failure to comply with these instructions may cause the warming stretcher to produce excessive temper-
atures resulting in thermal injury to the patient.
1. For easier servicing, raise the litter to the full up position and remove the mattress.
2. Using a standard screwdriver, carefully remove the two end caps (item AG, page 66) from the display as-
sembly.
3. Using pliers, remove the strain relief bushing (item K, page 66) and disconnect the cable from the display
board.
Summary of Contents for Thermaltek 1030
Page 8: ...Notes 5 2 ...
Page 15: ...Notes 10 2 ...
Page 29: ...24 715 1 250 Side Control Base Assembly with Brakes ...
Page 39: ...Notes 32 2 ...
Page 41: ...Side Control Base Assembly with Jacks 34 Assembly part number 715 1 260 ...
Page 45: ...Side Control Base Assembly With Jacks 37 1 Assembly part number 1210 2 10 ...
Page 48: ...Notes 37 4 ...
Page 49: ...End Control Base Assembly with Jacks 38 FOOT END HEAD END Assembly part number 716 1 270 ...
Page 56: ...43 Notes ...
Page 74: ...57 4 Notes ...
Page 91: ...74 Notes ...
Page 116: ...DH 2 97 1030 90 2 REV A 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...