![Stryker Medical Renaissance Series Maintenance Manual Download Page 22](http://html.mh-extra.com/html/stryker-medical/renaissance-series/renaissance-series_maintenance-manual_1386144022.webp)
Service Information
21
Displaying Individual Load Cell Outputs:
A faulty load cell can be detected by entering diagnostics and displaying individual load cell outputs.
1.
Enter the diagnostic mode. The scale monitor will display DOWNLOAD BARCODE when the diagnostic
mode is active.
2.
Press and release LBS/KGS until the scale monitor displays DISPLAY CORNER.
3.
Press and release WEIGHT SYSTEM ON. The scale monitor should display PICK CORNER NOW.
4.
Press and release the position key that corresponds with the load cell to be checked. The scale monitor
should display XX=NNN.N. ”XX” represents the initials of the selected corner, i.e. H/R will be displayed
for the patient’s head end, right side. ”NNN.N” represents the actual weight load on the load cell.
5.
Repeat step four for each corner. Head end weight readings will normally be lower than foot end weights.
Weight readings should be constant. A jumpy, drifting 000.0 or 999.9 weight indicates a problem with
the selected load cell assembly or load cell cable.
6.
When all the load cell outputs have been checked, press and release WEIGHT SYSTEM ON.
7.
Press and release LBS/KGS until the scale monitor displays QUIT/SAVE.
8.
Press and release WEIGHT SYSTEM ON. The scale monitor should shut off after one second.
LOAD CELL LOCATIONS
Summary of Contents for Renaissance Series
Page 33: ...Base Assembly 32 Assembly part number 2020 1 210 reference only ...
Page 37: ...Base Assembly with 110V Outlet Option 36 Assembly part number 2020 11 210 reference only ...
Page 39: ...Base Hydraulics Assembly 38 Assembly part number 2020 70 210 reference only ...
Page 40: ...Base Hydraulics Assembly 39 ...
Page 49: ...International Hood and Base Assembly 48 N FUSE JUNCTION BLOCK CORD ...
Page 52: ...Basic Litter Assembly 51 B Assembly part number 2020 1 513 reference only ...
Page 53: ...Basic Litter Assembly 52 LITTER CONTROL BOARD CONNECTIONS MOTOR WIRE ROUTING E ...
Page 64: ...Notes 63 ...
Page 115: ...2020 250 410 Integrated Dynamic Mattress System Assembly 114 ...
Page 123: ...DH 2 00 2020 1 152 REV K 6300 Sprinkle Road Kalamazoo MI 49001 ...