STEER CASTER LINKAGE ADJUSTMENT
Tools Required:
Phillips Head Screwdriver
3/8” Open End Wrench
1/2” Open End Wrench
5/32” Allen Wrench
F
E
B
C
A
D
E
Adjustment Procedure:
1.
Lift up and support base hood (see page 15).
2.
Using a 5/32” Allen wrench and 3/8” open end wrench, remove items (D), (E) and (F).
3.
Using a 1/2” open end wrench, loosen jam nut (C).
4.
Tighten item (A) into item (B) for more lever engagement or away from item (B) for less engagement.
5.
To reassemble linkage, reverse steps 2 and 3.
6.
Check steering wheel engagement. When properly adjusted, the locking tab should clear the notch in
the steer caster when in the neutral position, and should fully engage when in steer.
7.
After proper adjustment is determined, reinstall base hood.
Summary of Contents for Renaissance Series
Page 33: ...Base Assembly 32 Assembly part number 2020 1 210 reference only ...
Page 37: ...Base Assembly with 110V Outlet Option 36 Assembly part number 2020 11 210 reference only ...
Page 39: ...Base Hydraulics Assembly 38 Assembly part number 2020 70 210 reference only ...
Page 40: ...Base Hydraulics Assembly 39 ...
Page 49: ...International Hood and Base Assembly 48 N FUSE JUNCTION BLOCK CORD ...
Page 52: ...Basic Litter Assembly 51 B Assembly part number 2020 1 513 reference only ...
Page 53: ...Basic Litter Assembly 52 LITTER CONTROL BOARD CONNECTIONS MOTOR WIRE ROUTING E ...
Page 64: ...Notes 63 ...
Page 115: ...2020 250 410 Integrated Dynamic Mattress System Assembly 114 ...
Page 123: ...DH 2 00 2020 1 152 REV K 6300 Sprinkle Road Kalamazoo MI 49001 ...