background image

Operation Guide

8

USING THE OPTIONAL X–RAY CASSETTE HOLDER

C

B

D

A

B

C

1. To access the Fowler x–ray cassette holder, raise the Fowler section.

2. Grasp handles (A) and squeeze, allowing locating pins (B) to disengage from mounting brackets (C).

3. Lower tray and install x–ray cassette.

4. Reverse steps 1 & 2 to engage loaded tray to Fowler mounting brackets (C).

NOTE
The tray position can be adjusted, from the patient’s head to the buttocks area, by loosening knob (D), sliding
the tray to the desired position, and re–tightening (D) to hold the position.

Summary of Contents for 2010

Page 1: ...2010 Critical Care Bed OPERATIONS MANUAL For Parts or Technical Assistance 1 800 327 0770 ...

Page 2: ...older 5 Trendelenberg Fowler Display Option 6 Operating the Emergency Drop Fowler 6 Operating Optional Permanently Attached I V Pole 7 Using the Equipment Tray Options 7 Using the Optional X Ray Cassette Holder 8 Weigh System Control Panel Guide 9 10 Weigh System Usage 11 12 Warranty Obtaining Parts and Service 13 Supplemental Warranty Coverage 13 Return Authorization 14 Freight Damage Claims 14 ...

Page 3: ...endelenberg Reverse Trendelenberg 12 degrees to 12 degrees Weigh System Accuracy optional equipment 1 of total patient weight WARNING CAUTION NOTE DEFINITION The words WARNING CAUTION and NOTE carry special meanings and should be carefully reviewed WARNING The personal safety of the patient or user may be involved Disregarding this information could result in injury to the patient or user CAUTION ...

Page 4: ...s the pedal marked with the Trendelenberg symbol head down Assure the head end lowers to the full down position S Raise the bed to the full up position and depress the pedal marked with the reverse Trendelenberg symbol foot down Assure the foot end lowers to the full down position S Raise the bed to the full up position Pivot the Fowler crank up into position Turn the crank clock wise to raise the...

Page 5: ...ee Gatch Lift Pedal Pump pedal to raise litter Trendelenberg Pedal Depress pedal to lower head end Note Depress both Trendelenberg pedals at the same time to lower litter Reverse Trendelenberg Pedal Depress pedal to lower foot end Note Depress both Trendelenberg pedals at the same time to lower litter Brake Steer Pedal Depress pedal fully to set four wheel brakes Lift pedal fully to engage steer c...

Page 6: ...e steer caster For proper tracking of the steer caster push the bed approximately 10 feet to allow the wheels to face the direction of travel before engaging the steer pedal Without doing this proper tracking will not occur and it will be difficult to steer the bed OPERATING OPTIONAL OXYGEN BOTTLE HOLDER A 1 To use oxygen bottle holder rotate A upward until it stops in an upright position as shown...

Page 7: ...ed on the battery holder The batteries will need to be replaced when the Trend Fowler indicator lights appear dim OPERATING THE EMERGENCY DROP FOWLER FRAME END A B The Emergency Drop Fowler Release is located at the foot of the bed to the right of the Fowler crank To lower the Fowler quickly in an emergency S Press the button on the knob A and pull the knob out until it stops S The Fowler will low...

Page 8: ...rtion C Raise the pole to the desired height 3 Turn the lock actuator B clockwise to lock the telescoping portion in place CAUTION The weight of the I V bags should not exceed 40 pounds USING THE EQUIPMENT TRAY OPTIONS CAUTION Both the equipment defibrillator tray and the equipment tray footboard combination options have a 40 lb limit To use the equipment tray footboard combination option mount as...

Page 9: ...ueeze allowing locating pins B to disengage from mounting brackets C 3 Lower tray and install x ray cassette 4 Reverse steps 1 2 to engage loaded tray to Fowler mounting brackets C NOTE The tray position can be adjusted from the patient s head to the buttocks area by loosening knob D sliding the tray to the desired position and re tightening D to hold the position ...

Page 10: ...n the ZERO BED function Press and hold ZERO BED KEY NOW UNLOCKED HOLD TO ZERO WT RELEASE TO ZERO Release ZERO BED DO NOT TOUCH BED 0 0 LB After putting new patient on bed Press and release SCALE SYSTEM ON Press and release PATIENT WEIGHT XXX X LB Press and hold ZERO GAIN LOSS HOLD TO ZERO G L RELEASE TO ZERO Release ZERO GAIN LOSS DO NOT TOUCH BED XXX X LB G00 0 To add or remove equipment without ...

Page 11: ...ons can be locked out by repeating above steps If a button is pressed for a function that has been locked out the display will read KEY LOCKED OUT To unlock a function Press and release LOCK UNLOCK LOCK UNLOCK KEY Press and release selected button KEY NOW UNLOCKED To convert the patient s weight to WEIGHT NOW KGS kilograms press and release XXX X KG GXX X LBS KGS Repeat the procedure to return to ...

Page 12: ...OCK UNLOCK The scale monitor will read LOCK UNLOCK KEY S Press and hold ZERO BED The scale monitor will read KEY NOW UNLOCKED HOLD TO ZERO WT RELEASE TO ZERO S Release ZERO BED The scale monitor will now read DO NOT TOUCH BED 0 0 LB AFTER PUTTING A NEW PATIENT ON THE BED S Press and release SCALE SYSTEM ON to activate the weigh system The scale monitor will acti vate and immediately turn off S Aft...

Page 13: ...on again S To convert the patient s weight and weight gain loss to kilograms press and release SCALE SYS TEM ON to activate the weigh system After the scale monitor activates and then turns off press and release PATIENT WEIGHT After the scale monitor reads XXX X LB GXX X press and re lease the LBS KGS button The scale monitor will read WEIGHT NOW KGS XXX X KG GXX X S Repeat the procedure to return...

Page 14: ...OR CONSEQUENTIAL DAMAGES ARISING FROM OR IN ANY MANNER RELATED TO SALES OR USE OF ANY SUCH EQUIPMENT To Obtain Parts and Service Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives These representatives are factory trained available locally and carry a substantial spare parts inventory to minimize repair time Simply call your local representati...

Page 15: ...N THE DELIVERY RECEIPT AT THE TIME OF RECEIPT Upon prompt notification Stryker will file a freight claim with the appropriate carrier for damages incurred Claim will be limited in amount to the actual replacement cost In the event that this information is not received by Stryker within the fifteen 15 day period following the delivery of the merchandise or the damage was not noted on the delivery r...

Page 16: ...DH 6 92 2010 1 110 REV__ 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...