Operation Guide
6
TRENDELENBERG/FOWLER DISPLAY OPTION
The Trend/Fowler display option indicator lights are powered by 3 ”D” size (1.5 V each) alkaline batteries.
The batteries are located under the foot end frame end. To replace the batteries, remove the two screws on
the frame end bumper and lift off the plastic frame end. Disconnect the power cable and replace the batteries.
NOTE
Polarity of battery placement is marked on the battery holder.
The batteries will need to be replaced when the Trend/Fowler indicator lights appear dim.
OPERATING THE EMERGENCY DROP FOWLER
FRAME END
A
B
The Emergency Drop Fowler Release is located at the foot of the bed to the right of the Fowler crank. To
lower the Fowler quickly in an emergency:
S
Press the button on the knob (A) and pull the knob out until it stops.
S
The Fowler will lower until it is flat.
Resetting the Fowler:
S
Turn the Fowler crank (B) clockwise until it stops.
S
Push the button on the knob and push the knob in until it stops.
NOTE
The Fowler must be reset before it will raise again.