![Stryker Medical 1020 Maintenance Manual Download Page 16](http://html.mh-extra.com/html/stryker-medical/1020/1020_maintenance-manual_1386128016.webp)
Service Information
15
HYDRAULIC CHECK VALVE REPLACEMENT
Required Tools:
3/8 Open End Wrench
Stiff Wire (with bent, pointed end) Small Needle Nose Pliers
3/4 Open End Wrench
Torque Wrench (with Ft. Lbs. adjust.)
7/32 Hex Allen Wrench
1/2 Inch Diameter Rod
Replacement of Valve #1
WARNING
To avoid personal injury or damage to the stretcher, remove the litter and the base hood before beginning
service on the stretcher.
1. Using a 3/8 open end wrench, remove square head set screws from both head and foot end jack support
tubes. Remove litter top and set aside.
2. Lift base hood off base frame and set aside.
3. Lower the jack to full down position. The actuator must be manually lowered while depressing the appropri-
ate release pedal.
4. Remove the pin body assembly (4) with a 3/4 open end wrench and discard the housing gasket (5).
NOTE
Although the hydraulic fluid is not under pressure, some fluid loss will occur. The fluid loss should be minimal
but covering the floor is advisable.
5. Using a 7/32 hex Allen wrench, remove the valve plug (6).
6. Using a stiff wire with a bent, pointed end, remove and discard the valve (1) and the seal (7).
7. Install the new seal (7) flat to the bottom of its hole with a 1/2 inch diameter rod and install the new valve
(1) with the beveled end out (as shown in the illustration).
8. Install the valve plug (6) with the countersunk end first and the beveled end out. Tighten to 10 foot pounds
torque.
9. Install the pin body assembly (4) with the new housing gasket (5) and tighten to 10 foot pounds torque.
10. Pump up the jack to the maximum height. Apply weight to be sure the jack holds its position and there
are no hydraulic leaks before replacing the base hood and the litter.
ITEM
PART NO.
PART NAME
1
926–20–153
Check Valve
2
926–20–153
Check Valve
3
715–1–341
Poppet
4
715–100–312
Pin Housing Assembly*
715–270–100
Valve Assembly**
5
715–1–330
Housing Gasket
6
715–1–309
Valve Plug
7
926–20–154
Seal
8
715–1–101
Base Plug
9
926–20–156
Seal
10
715–1–309
Valve Plug
11
926–20–154
Seal
12
715–1–301
Base Plug
13
926–20–156
Seal
14
390–2–134
Conical Comp. Spring
*
Used on jack part number 715–100–310.
**
Used on jack part number 715–270–10.
(see label on side of jack reservoir for jack part number)
FILLER PLUG
3
14
12
13
Summary of Contents for 1020
Page 19: ...18 715 1 250 Side Control Base Assembly with Brakes ...
Page 30: ...Notes 29 ...
Page 33: ...Side Control Base Assembly with Jacks 32 Assembly part number 715 1 260 ...
Page 37: ...Side Control Base Assembly with Jacks 36 Assembly part number 1210 2 10 ...
Page 40: ...Notes 39 ...
Page 41: ...End Control Base Assembly with Jacks 40 FOOT END HEAD END Assembly part number 716 1 270 ...
Page 50: ...49 1020 1 10 Litter Assembly ...
Page 51: ...50 Litter Assembly ...
Page 65: ...64 1020 22 10 X Ray Tray Assembly Right Side Loading ...
Page 67: ...66 1020 21 10 X Ray Tray Assembly Left Side Loading AF BL BR BP ...
Page 93: ...DH 11 97 1020 32 34 REV K 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...