Service Information
10
CASTER COVER INSTALLATION AND REMOVAL
Double Prongs
Single Prong
align cover with axle nut or bolt head, as shown.
Push down on the opposite side of the cover until
single prong engages with caster horn.
Push with palm on cover until
Properly Attached
Cover
To remove wheel cover, insert large screwdriver into cut–out in
side of wheel cover and into the space between the double prongs.
Pry up cover to disengage double prongs and push sharply upward
double prongs engage.
1.
2.
3.
Looking through the larger of the two side cut–outs,
Top View (Cut–Away)
Top View (Cut–Away)
Top View (Cut–Away)
to disengage single prong.