CASTER MAINTENANCE
Required Tools:
5/8” Wrench
11/16” Wrench
Maintenance Procedure:
A
B
C
D
1. Remove the plastic caster cover (see page 10).
2. Using the 5/8” wrench and the 11/16” wrench, remove the centerlock nut (item A) from the through bolt
(item B) for the caster wheel.
3. Support the corner of the stretcher where the wheel is being removed and remove the through bolt (item
B) and the molded wheel (item C) .
4. Clean the through bolt, molded wheel, and the inside of the caster horn (item D) removing any dirt and
debris. Ensure the bearings in the molded wheel spin freely and easily.
5. Replace the molded wheel and the through bolt.
6. Replace the centerlock nut on the through bolt and use the 5/8” and 11/16” wrenches to tighten it securely.
* See page 27 for replacement part numbers.
Summary of Contents for 1020
Page 19: ...18 715 1 250 Side Control Base Assembly with Brakes ...
Page 30: ...Notes 29 ...
Page 33: ...Side Control Base Assembly with Jacks 32 Assembly part number 715 1 260 ...
Page 37: ...Side Control Base Assembly with Jacks 36 Assembly part number 1210 2 10 ...
Page 40: ...Notes 39 ...
Page 41: ...End Control Base Assembly with Jacks 40 FOOT END HEAD END Assembly part number 716 1 270 ...
Page 50: ...49 1020 1 10 Litter Assembly ...
Page 51: ...50 Litter Assembly ...
Page 65: ...64 1020 22 10 X Ray Tray Assembly Right Side Loading ...
Page 67: ...66 1020 21 10 X Ray Tray Assembly Left Side Loading AF BL BR BP ...
Page 93: ...DH 11 97 1020 32 34 REV K 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...