Error Message
Explanation
Action
26
Information
Specimen mover plate not
lowered!
The pneumatic head with
specimen mover plate is not
down when the process is
started.
Appears if method is started
in single specimen mode
(SS) and mover plate is not
lowered.
Lower the pneumatic head.
Re-start the process.
9
Technical data
9.1
Technical data - LaboForce-100
Subject
Specifications
Motor output
180 W
Rotational speed
50-150 rpm
Rotational direction
Clockwise/Counter-clockwise
Force, individual
specimens
10-50 N
Force, specimen holder
30-300 N
Safety standards
See the Declaration of Conformity
REACH
For information about REACH, contact your local Struers office.
Noise level
See the noise level stated for: LaboPol
Power supply
LaboForce-100 is connected directly to LaboPol
Software and electronics
Controls
Touch pad, Turn/push knob knob
Memory
FLASH-ROM/RAM/NV-RAM
LC display
TFT-colour 320x240 dots with LED
back light
Operating environment
Surrounding
temperature
5 – 40°C/41 – 104°F
Humidity
< 85 % RH non-condensing
Storage and transport
conditions
Surrounding
temperature
0 – 60°C/32 – 140°F
Humidity
< 85 % RH non-condensing
28
LaboForce-100
9 Technical data